EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC 4.0

Optimization of Leukocytic Flags in the Sysmex‐ XN Series With the Help of Flow Cytometric Differential ( FlowDiff ) Can Significantly Optimize Laboratory Efficiency

Ninna Hahn Tougaard iD, Snorre Kristian Vesten Pedersen, Anders Abildgaard iD, Anne Winther-Larsen iD, Konstantinos Dimopoulos iD

International Journal of Laboratory Hematology · 03.11.2026

DOI: 10.1111/ijlh.70275 ↗

Kurz gefasst

Fachartikel in der Zeitschrift International Journal of Laboratory Hematology, online erschienen am 03.11.2026. Verfasst von Ninna Hahn Tougaard und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Biochemistry, Bispebjerg and Frederiksberg Hospital, Copenhagen, Denmark. Die Arbeit nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

The aim of this study was to adjust the default white blood cell (WBC) differential count flag settings on the Sysmex XN‐9100 haematology analyser to reduce unnecessary additional leukocyte differential counts without compromising diagnostic sensitivity.

Methods

Data from 4554 routine blood samples flagged on Sysmex XN‐9100 and corresponding additional leukocyte differential counts performed on the Sysmex XF‐1600 flow cytometer were used to optimise the WBC flagging rules.

Results

A total of 1148 samples had scattergram‐related flags, which are non‐adjustable. Optimal cut points for Q‐flags were established based on Youden's Index for the remaining 3406 samples, while the diagnostic sensitivity of the new flagging rules for acute leukaemia was assessed in a validation cohort. The optimal cut points for Q‐flags were 120 for ‘Atypical Lymphocytes’, 170 for ‘Abnormal Lymphocytes’ and 160 for ‘Blasts’, while the Q‐flag ‘Left shift’ offered no diagnostic value in our cohort. The count flags ‘eosinophilia’, ‘basophilia’ and ‘monocytopenia’ were deactivated, as the results from Sysmex XN were comparable to the flow cytometric reference method. By implementing the new flagging rules, 1709 (37.5%) of the samples would have been processed without need for additional leukocyte differential count. When evaluated in a separate validation cohort of 90 samples from patients with newly diagnosed acute leukaemia, the new rules showed similar diagnostic sensitivity to the default flagging system.

Conclusion

Optimisation of Sysmex flagging rules for leukocyte differential counts was feasible and significantly reduced unnecessary reflex testing, while maintaining analytical accuracy and diagnostic safety.

Schlagwörter

Stichwörter der Autor:innen

FlowDiffdifferential countflagshaematoanalyserhaematology

Autor:innen und Institutionen

  • Department of Clinical Biochemistry, Bispebjerg and Frederiksberg Hospital, Copenhagen, Denmark.
  • Department of Clinical Biochemistry, Aarhus University Hospital, Aarhus, Denmark.
  • Department of Clinical Medicine, Aarhus University, Aarhus, Denmark.
  • Department of Clinical Medicine, Faculty of Health and Medical Sciences, University of Copenhagen, Copenhagen, Denmark.

ORCID-Profile: Ninna Hahn Tougaard, Anders Abildgaard, Anne Winther-Larsen, Konstantinos Dimopoulos

Zitieren

APA

Tougaard, N. H., Pedersen, S. K. V., Abildgaard, A., Winther-Larsen, A., & Dimopoulos, K. (2026). Optimization of Leukocytic Flags in the Sysmex‐ XN Series With the Help of Flow Cytometric Differential ( FlowDiff ) Can Significantly Optimize Laboratory Efficiency. International Journal of Laboratory Hematology. https://doi.org/10.1111/ijlh.70275

Vancouver

Tougaard NH, Pedersen SKV, Abildgaard A, Winther-Larsen A, Dimopoulos K. Optimization of Leukocytic Flags in the Sysmex‐ XN Series With the Help of Flow Cytometric Differential ( FlowDiff ) Can Significantly Optimize Laboratory Efficiency. International Journal of Laboratory Hematology. 2026. doi:10.1111/ijlh.70275

BibTeX
@article{Tougaard2026,
  author = {Tougaard, Ninna Hahn and Pedersen, Snorre Kristian Vesten and Abildgaard, Anders and Winther-Larsen, Anne and Dimopoulos, Konstantinos},
  title = {Optimization of Leukocytic Flags in the Sysmex‐ XN Series With the Help of Flow Cytometric Differential ( FlowDiff ) Can Significantly Optimize Laboratory Efficiency},
  journal = {International Journal of Laboratory Hematology},
  year = {2026},
  doi = {10.1111/ijlh.70275}
}
Kostenlos starten →