EvaluationsstudieOpen Access

Evaluation of the BT ‐50 Automatic Quality Control Unit on the XN9000 Hematology Analyzer

Congyu Zou iD, Yan Wei, Yuyin Liu, Xiaoqian Zeng iD

International Journal of Laboratory Hematology · 2026

DOI: 10.1111/ijlh.70184 ↗

Kurz gefasst

Evaluationsstudie in der Zeitschrift International Journal of Laboratory Hematology, 2026. Verfasst von Congyu Zou und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Laboratory, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, China. Hauptthemen laut MEDLINE: Quality Control, Hematology, Automation, Laboratory. Die Arbeit nennt 12 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

This study aims to explore the application of BT50‐based automatic quality control in the XN9000 hematology analyzer line, comparing its benefits and limitations with manual QC procedures, and to identify potential improvements, thereby providing a reference for the development of automated QC in clinical laboratories.

Methods

Both manual and automatic QC procedures were performed for blood cell analysis. Manual QC data (January 1, 2024, to July 31, 2024) and automatic QC data (August 1, 2024, to February 28, 2025) were collected from the Department of Clinical Laboratory, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology. SPSS 18.0 software was used for the statistical analysis. The parameters analyzed included the coefficient of variation (CV), daily QC completion time, first sample review time, and staff workload.

Results

(1) Compared to manual QC, no statistically significant differences were observed in the CV values of any parameters between the two methods (Mann–Whitney U test, p > 0.05 for all). (2) The median completion time for automatic QC was 42 min earlier than that for manual QC, with automatic QC finishing at 07:48:47 (IQR: 07:47:55, 07:48:46) and manual QC finishing at 08:30:34 (IQR: 08:24:17, 08:37:32). (3) With a daily work start time of 08:00, the median first sample review time was advanced by 20 min under automatic QC, with manual QC reviewed at 08:48:00 (IQR: 08:36:00, 08:50:30) and automatic QC reviewed at 08:28:00 (IQR: 08:21:30, 09:07:30). (4) The automatic QC system significantly reduced the number of manual steps, including those for storage, rewarming, mixing, and post‐analysis storage.

Conclusion

Implementing the automatic QC system facilitates earlier report review, reduces workload, enhances operational efficiency, and maintains CV values within acceptable laboratory standards.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Quality ControlHematologyAutomation, LaboratoryHumansLaboratories, ClinicalBlood Cell Count

Stichwörter der Autor:innen

automatic quality controlcoefficient of variation (CV)hematology analysismanual quality control

Autor:innen und Institutionen

  • Department of Clinical Laboratory, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, China.

ORCID-Profile: Congyu Zou, Xiaoqian Zeng

Zitieren

APA

Zou, C., Wei, Y., Liu, Y., & Zeng, X. (2026). Evaluation of the BT ‐50 Automatic Quality Control Unit on the XN9000 Hematology Analyzer. International Journal of Laboratory Hematology. https://doi.org/10.1111/ijlh.70184

Vancouver

Zou C, Wei Y, Liu Y, Zeng X. Evaluation of the BT ‐50 Automatic Quality Control Unit on the XN9000 Hematology Analyzer. International Journal of Laboratory Hematology. 2026. doi:10.1111/ijlh.70184

BibTeX
@article{Zou2026,
  author = {Zou, Congyu and Wei, Yan and Liu, Yuyin and Zeng, Xiaoqian},
  title = {Evaluation of the BT ‐50 Automatic Quality Control Unit on the XN9000 Hematology Analyzer},
  journal = {International Journal of Laboratory Hematology},
  year = {2026},
  doi = {10.1111/ijlh.70184}
}
Kostenlos starten →