EUVIMEDEUVIMED entdecken
Open Access

Refining quality control strategies in highly automated laboratories

María Costa-Pallaruelo, Álvaro García-Osuna, Marina Canyelles, Cecília Martínez-Bru, Nicoleta Nan, Rosa Ferrer-Perez, Francisco Blanco-Vaca, Leonor Guiñón

Biochemia medica · 2025

DOI: 10.11613/bm.2025.030704 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biochemia medica, 2025. Verfasst von María Costa-Pallaruelo und 7 weiteren Autor:innen. Beteiligte Einrichtung: Biochemistry Department, Hospital de la Santa Creu i Sant Pau, IIB Sant Pau, Barcelona, Spain. Hauptthemen laut MEDLINE: Risk Management, Automation, Laboratory, Laboratories, Clinical. Die Arbeit nennt 23 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

The ISO 15189:2022 standard considers both the robustness of analytical methods and the risk of erroneous results in the quality control plan (QCP) design. Westgard et al.’s nomogram recommends quality control (QC) rules based on sample run size to ensure that the maximum expected number of unreliable patient results remains below one. This study aimed to implement a standardized, risk-based QC strategy across multiple analyzers without integrated on board QC, ensuring practical quality assurance. Material and methods: Thirty-two biochemistry parameters on Alinity c systems and three on Cobas Pro systems were included. The analytical performance of each parameter on each analyzer was assessed using sigma metric. Workload requirements were considered to determine the desired run size. Based on the “sigma metric statistical QC run size nomogram” proposed by Westgard et al., a multistage bracketed QCP was designed for each parameter. When multiple designs were available, the simplest QC rule was prioritized.

Results

Seven QCPs were initially established for 35 parameters. In the absence of automation, practical adaptations based on sigma metrics were implemented. Additionally, to streamline management, the QCP covering the greatest number of parameters per analyzer was prioritized, which ultimately resulted in the adoption of only two general QCP. Only 4 individualized QCP were required to cover 10 parameters with lower sigma values.

Conclusions

This approach demonstrates the feasibility of implementing a refined QC strategy for parameters with sigma ≥ 4 in a highly automated laboratory, ensuring consistent quality assurance and efficient resource allocation for higher-risk tests.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Risk ManagementAutomation, LaboratoryLaboratories, ClinicalQuality ControlHumansBiochemistry

Stichwörter der Autor:innen

Max E(Nuf) QC modellaboratory automationmultistage QCrisk managementrun size

Autor:innen und Institutionen

  • Biochemistry Department, Hospital de la Santa Creu i Sant Pau, IIB Sant Pau, Barcelona, Spain.
  • Core Laboratory, Hospital de la Santa Creu i Sant Pau, Barcelona, Spain.
  • Quality Department Laboratories, Hospital de la Santa Creu i Sant Pau, Barcelona, Spain.

Zitieren

APA

Costa-Pallaruelo, M., García-Osuna, Á., Canyelles, M., Martínez-Bru, C., Nan, N., Ferrer-Perez, R., Blanco-Vaca, F., & Guiñón, L. (2025). Refining quality control strategies in highly automated laboratories. Biochemia medica. https://doi.org/10.11613/bm.2025.030704

Vancouver

Costa-Pallaruelo M, García-Osuna Á, Canyelles M, Martínez-Bru C, Nan N, Ferrer-Perez R, et al. Refining quality control strategies in highly automated laboratories. Biochemia medica. 2025. doi:10.11613/bm.2025.030704

BibTeX
@article{CostaPallaruelo2025,
  author = {Costa-Pallaruelo, María and García-Osuna, Álvaro and Canyelles, Marina and Martínez-Bru, Cecília and Nan, Nicoleta and Ferrer-Perez, Rosa and Blanco-Vaca, Francisco and Guiñón, Leonor},
  title = {Refining quality control strategies in highly automated laboratories},
  journal = {Biochemia medica},
  year = {2025},
  doi = {10.11613/bm.2025.030704}
}