Robustness assessment of an automated AI ‐based white blood cell morphometric analysis system using different smear preparation methods
International Journal of Laboratory Hematology · 2024
DOI: 10.1111/ijlh.14350 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Laboratory Hematology, 2024. Verfasst von Mendamar Ravzanaadii und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Laboratory Medicine, Juntendo University Graduate School of Medicine, Tokyo, Japan. Hauptthemen laut MEDLINE: Leukocytes. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 24 Literaturangaben. Gefördert von: Grants-in-Aid for Scientific Research, International Joint Research Programs from the Japan Society for the Promotion of Science.
Abstract (Originalsprache)
Introduction
Numerous AI‐based systems are being developed to evaluate peripheral blood (PB) smears, but the feasibility of these systems on different smear preparation methods has not been fully understood. In this study, we assessed the impact of different smear preparation methods on the robustness of the deep learning system (DLS).
Methods
We collected 193 PB samples from patients, preparing a pair of smears for each sample using two systems: (1) SP50 smears, prepared by the DLS recommended fully automated slide preparation with double fan drying and staining (May–Grunwald Giemsa, M–G) system using SP50 (Sysmex) and (2) SP1000i smears, prepared by automated smear preparation with single fan drying by SP1000i (Sysmex) and manually stained with M–G. Digital images of PB cells were captured using DI‐60 (Sysmex), and the DLS performed cell classification. Sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) were used to evaluate the performance of the DLS.
Results
The specificity and NPV for all cell types were 97.4%–100% in both smear sets. The average sensitivity and PPV were 88.9% and 90.1% on SP50 smears, and 87.0% and 83.2% on SP1000i smears, respectively. The lower performance on SP1000i smears was attributed to the intra‐lineage misclassification of neutrophil precursors and inter‐lineage misclassification of lymphocytes.
Conclusion
The DLS demonstrated consistent performance in specificity and NPV for smears prepared by a system different from the recommended method. Our results suggest that applying an automated smear preparation system optimized for the DLS system may be important.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Clinical Laboratory Medicine, Juntendo University Graduate School of Medicine, Tokyo, Japan.
- Department of Advanced Research Institute for Health Science, Juntendo University Graduate School of Medicine, Tokyo, Japan.
- Scientific Affairs, Sysmex Corporation, Kobe, Japan.
- Sysmex Corporation, Kobe, Japan.
- Division of Cell Therapy & Blood Transfusion Medicine, Juntendo University School of Medicine, Tokyo, Japan.
- Department of Hematology, Juntendo University Graduate School of Medicine, Tokyo, Japan.
ORCID-Profile: Mendamar Ravzanaadii, Yuki Horiuchi
Förderung
- Grants-in-Aid for Scientific Research
- International Joint Research Programs from the Japan Society for the Promotion of Science
Zitieren
APA
Ravzanaadii, M., Horiuchi, Y., Iwasaki, Y., Matsuzaki, A., Kaniyu, K., Bai, J., Konishi, A., Ando, J., Ando, M., & Tabe, Y. (2024). Robustness assessment of an automated AI ‐based white blood cell morphometric analysis system using different smear preparation methods. International Journal of Laboratory Hematology. https://doi.org/10.1111/ijlh.14350
Vancouver
Ravzanaadii M, Horiuchi Y, Iwasaki Y, Matsuzaki A, Kaniyu K, Bai J, et al. Robustness assessment of an automated AI ‐based white blood cell morphometric analysis system using different smear preparation methods. International Journal of Laboratory Hematology. 2024. doi:10.1111/ijlh.14350
BibTeX
@article{Ravzanaadii2024,
author = {Ravzanaadii, Mendamar and Horiuchi, Yuki and Iwasaki, Yosuke and Matsuzaki, Akihiko and Kaniyu, Kimiko and Bai, Jing and Konishi, Aya and Ando, Jun and Ando, Miki and Tabe, Yoko},
title = {Robustness assessment of an automated AI ‐based white blood cell morphometric analysis system using different smear preparation methods},
journal = {International Journal of Laboratory Hematology},
year = {2024},
doi = {10.1111/ijlh.14350}
}
EUVIMED