Vergleichsstudie

Performance comparison of two automated digital morphology analyzers for leukocyte differential in patients with malignant hematological diseases: Mindray MC ‐80 and Sysmex DI ‐60

Xianfei Ye iD, Lijuan Fang, Yunying Chen, Jixiang Tong, Xiaoni Ning, Lanjun Feng, Yuting Xu, Dagan Yang iD

International Journal of Laboratory Hematology · 2024

DOI: 10.1111/ijlh.14227 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift International Journal of Laboratory Hematology, 2024. Verfasst von Xianfei Ye und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratory Medicine, the First Affiliated Hospital, Zhejiang University School of Medicine, Hangzhou, People's Republic of China. Hauptthemen laut MEDLINE: Leukocytes. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 25 Literaturangaben. Gefördert von: Scientific and Technological Innovation 2030-"New Generation of Artificial Intelligence" Major Project.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The MC‐80 (Mindray, Shenzhen, China), a newly available artificial intelligence (AI)‐based digital morphology analyzer, is the focus of this study. We aim to compare the leukocyte differential performance of the Mindray MC‐80 with that of the Sysmex DI‐60 and the gold standard, manual microscopy.

Methods

A total of 100 abnormal peripheral blood (PB) smears were compared across the MC‐80, DI‐60, and manual microscopy. Sensitivity, specificity, predictive value, and efficiency were calculated according to the Clinical and Laboratory Standards Institute (CLSI) EP12‐A2 guidelines. Comparisons were made using Bland–Altman analysis and Passing‐Bablok regression analysis. Additionally, within‐run imprecision was evaluated using five samples, each with varying percentages of mature leukocytes and blasts, in accordance with CLSI EP05‐A3 guidelines.

Results

The within‐run coefficient of variation (%CV) of the MC‐80 for most cell classes in the five samples was lower than that of the DI‐60. Sensitivities for the MC‐80 ranged from 98.2% for nucleated red blood cells (NRBC) to 28.6% for reactive lymphocytes. The DI‐60's sensitivities varied between 100% for basophils and reactive lymphocytes, and 11.1% for metamyelocytes. Both analyzers demonstrated high specificity, negative predictive value, and efficiency, with over 90% for most cell classes. However, the DI‐60 showed relatively lower specificity for lymphocytes (73.2%) and lower efficiency for blasts and lymphocytes (80.1% and 78.6%, respectively) compared with the MC‐80. Bland–Altman analysis indicated that the absolute mean differences (%) ranged from 0.01 to 4.57 in MC‐80 versus manual differential and 0.01 to 3.39 in DI‐60 versus manual differential. After verification by technicians, both analyzers exhibited a very high correlation ( r = 0.90–1.00) with the manual differential results in neutrophils, lymphocytes, and blasts.

Conclusions

The Mindray MC‐80 demonstrated good performance for leukocyte differential in PB smears, notably exhibiting higher sensitivity for blasts identification than the DI‐60.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

LeukocytesHumansSensitivity and SpecificityHematologic NeoplasmsLeukocyte CountFemaleAutomation, LaboratoryMaleReproducibility of ResultsArtificial Intelligence

Stichwörter der Autor:innen

Mindray MC‐80Sysmex DI‐60digital morphology analyzerperformance

Autor:innen und Institutionen

  • Department of Laboratory Medicine, the First Affiliated Hospital, Zhejiang University School of Medicine, Hangzhou, People's Republic of China.
  • Key Laboratory of Clinical In Vitro Diagnostic Techniques of Zhejiang Province, Hangzhou, People's Republic of China.
  • Hangzhou Dian Medical Laboratory Center Co., Ltd, Hangzhou, People's Republic of China.
  • Department of Laboratory Medicine, Hangzhou Children's Hospital, Hangzhou, People's Republic of China.
  • Department of Hematology, the First Affiliated Hospital, Zhejiang University School of Medicine, Hangzhou, People's Republic of China.

ORCID-Profile: Xianfei Ye, Dagan Yang

Förderung

  • Scientific and Technological Innovation 2030-"New Generation of Artificial Intelligence" Major Project

Zitieren

APA

Ye, X., Fang, L., Chen, Y., Tong, J., Ning, X., Feng, L., Xu, Y., & Yang, D. (2024). Performance comparison of two automated digital morphology analyzers for leukocyte differential in patients with malignant hematological diseases: Mindray MC ‐80 and Sysmex DI ‐60. International Journal of Laboratory Hematology. https://doi.org/10.1111/ijlh.14227

Vancouver

Ye X, Fang L, Chen Y, Tong J, Ning X, Feng L, et al. Performance comparison of two automated digital morphology analyzers for leukocyte differential in patients with malignant hematological diseases: Mindray MC ‐80 and Sysmex DI ‐60. International Journal of Laboratory Hematology. 2024. doi:10.1111/ijlh.14227

BibTeX
@article{Ye2024,
  author = {Ye, Xianfei and Fang, Lijuan and Chen, Yunying and Tong, Jixiang and Ning, Xiaoni and Feng, Lanjun and Xu, Yuting and Yang, Dagan},
  title = {Performance comparison of two automated digital morphology analyzers for leukocyte differential in patients with malignant hematological diseases: Mindray MC ‐80 and Sysmex DI ‐60},
  journal = {International Journal of Laboratory Hematology},
  year = {2024},
  doi = {10.1111/ijlh.14227}
}
Kostenlos starten →