EUVIMEDOpdag EUVIMED
Crossref

Analytical validation of white blood cell differential and platelet assessment on the Sysmex DI ‐60 digital morphology analyzer

Ivana Lapić iD, Marija Miloš iD, Marija Dorotić, Valentina Drenški, Désirée Coen Herak, Dunja Rogić

International Journal of Laboratory Hematology · 2023

DOI: 10.1111/ijlh.14101 ↗

At a glance

Journal article in International Journal of Laboratory Hematology, 2023. Written by Ivana Lapić and 5 co-authors. Institution: Department of Laboratory Diagnostics, University Hospital Center Zagreb, Zagreb, Croatia. Major MEDLINE topics: Leukocytes, Lymphocytes. Cited 14 times so far (Crossref); 19 references.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Introduction

Digital morphology analyzers are increasingly replacing light microscopy in laboratory hematology practice. This study aimed to perform the analytical validation of the white blood cell (WBC) differential and of reliability of platelet assessment on Sysmex DI‐60 (Kobe, Japan).

Methods

Validation included determination of within‐run and between‐run precision for WBC differential according to the CLSI EP15‐A3 protocol, accuracy and method comparison with light microscopy and with the automated WBC differential from the Sysmex XN‐10 hematology analyzer, reliability of platelet clump detection and platelet count estimation.

Results

Standard deviations of both pre‐ and post‐classification mostly satisfied manufacturer's criteria for imprecision. Accuracy assessment revealed that only eosinophil count (1.4%) in one peripheral blood smear (PBS) remained outside the declared range (2–10%) after reclassification. Method comparison between DI‐60 and light microscopy yielded Spearman's correlation coefficients from 0.37 (basophils) to 0.94 (neutrophils and lymphocytes), minor proportional difference for bands, constant difference for monocytes, both constant and proportional difference for lymphocytes and statistically significant biases for bands, lymphocytes, monocytes and basophils. Diagnostic sensitivity (Se) and specificity (Sp) of DI‐60 in detecting immature/pathological cells were 88.7% (95%CI:81.1–94.0) and 83.0% (95%CI:78.7–86.7), respectively, with the area under the curve (AUC) of 0.86 (95%CI:0.82–0.89). Agreement in detection of platelet clumps was 94.8% (kappa coefficient = 0.67, 95%CI:0.53–0.80). Se and Sp of DI‐60 to detect platelet clumps were 65.7% (95%CI: 47.8–80.9) and 96.9% (95%CI: 93.9–98.6), respectively, while AUC was 0.81 (95%CI: 0.76–0.86).

Conclusion

DI‐60 provides reliable WBC differential and platelet assessment. In doubtful cases, the use of light microscopy is still mandatory.

Subjects

MeSH terms (MEDLINE)

LeukocytesLymphocytesHumansReproducibility of ResultsLeukocyte CountMonocytes

Author keywords

automationleukocytesmorphological and microscopic findingsplateletsvalidation study

Authors and institutions

  • Department of Laboratory Diagnostics, University Hospital Center Zagreb, Zagreb, Croatia.
  • Faculty of Pharmacy, University of Mostar, Mostar, Bosnia and Herzegovina.
  • Department of Medical Biochemistry and Laboratory Medicine, Merkur University Hospital, Zagreb, Croatia.
  • Faculty of Pharmacy and Biochemistry, University of Zagreb, Zagreb, Croatia.

ORCID profiles: Ivana Lapić, Marija Miloš

Cite

APA

Lapić, I., Miloš, M., Dorotić, M., Drenški, V., Coen Herak, D., & Rogić, D. (2023). Analytical validation of white blood cell differential and platelet assessment on the Sysmex DI ‐60 digital morphology analyzer. International Journal of Laboratory Hematology. https://doi.org/10.1111/ijlh.14101

Vancouver

Lapić I, Miloš M, Dorotić M, Drenški V, Coen Herak D, Rogić D. Analytical validation of white blood cell differential and platelet assessment on the Sysmex DI ‐60 digital morphology analyzer. International Journal of Laboratory Hematology. 2023. doi:10.1111/ijlh.14101

BibTeX
@article{Lapi2023,
  author = {Lapić, Ivana and Miloš, Marija and Dorotić, Marija and Drenški, Valentina and Coen Herak, Désirée and Rogić, Dunja},
  title = {Analytical validation of white blood cell differential and platelet assessment on the Sysmex DI ‐60 digital morphology analyzer},
  journal = {International Journal of Laboratory Hematology},
  year = {2023},
  doi = {10.1111/ijlh.14101}
}