EUVIMEDEUVIMED entdecken
Vergleichsstudie

Agreement Between Echocardiography and Ultrasonic Cardiac Output Monitoring (USCOM) for Hemodynamic Assessment in Pediatric Patients: A Bland–Altman Analysis

Saviga Sethasathien, Orana Arsasuwan, Theerapon Jariyasakoolroj, Suchaya Silvilairat

Echocardiography · 2026

DOI: 10.1111/echo.70645 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Echocardiography, 2026. Verfasst von Saviga Sethasathien und 3 weiteren Autor:innen. Beteiligte Einrichtung: Division of Pediatric Cardiology, Department of Pediatrics, Faculty of Medicine, Chiang Mai University, Chiang Mai, Thailand. Hauptthemen laut MEDLINE: Cardiac Output, Echocardiography, Heart Defects, Congenital, Echocardiography, Doppler. Die Arbeit nennt 19 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Echocardiography is widely used for non‐invasive measurements of stroke volume and cardiac output, while Ultrasonic Cardiac Output Monitor (USCOM) offers an alternative Doppler‐based approach. Whether these two modalities can be used interchangeably in clinical practice remains unclear. This study aims to evaluate the correlation and agreement between echocardiography and USCOM for hemodynamic assessment.

Methods

A total of 44 pediatric patients underwent hemodynamic assessment using echocardiography and sequential USCOM measurement during the same evaluation. Spearman's rank correlation was used to assess the relationship between methods. Agreement was evaluated using Bland‐Altman analysis.

Results

Strong correlations were found for stroke volume (SV; r = 0.97), cardiac output (CO; r = 0.95), cardiac index (CI; r = 0.91), stroke volume index (SVI; r = 0.90), and aortic valve (AV) diameter ( r = 0.91; p < 0.001), with a moderate correlation for left ventricular outflow tract (LVOT) velocity ( r = 0.64, p < 0.001). Bland‐Altman analysis revealed echocardiography produced lower values of all parameters compared to USCOM. Significant proportional bias was identified for SV, SVI, CO, and CI (all p ≤ 0.006), indicating that two methods increased disagreement with higher measured values. AV diameter and LVOT velocity showed no significant proportional bias and narrower limits of agreement.

Conclusions

Although echocardiography and USCOM correlate strongly for hemodynamic parameters, clinically meaningful systematic bias and proportional error limit their interchangeability for volumetric flow measurements. These devices should not be used interchangeably for quantitative hemodynamic monitoring, particularly in high‐output states.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cardiac OutputEchocardiographyHeart Defects, CongenitalEchocardiography, DopplerHumansFemaleReproducibility of ResultsMaleChildChild, PreschoolSensitivity and SpecificityInfantHemodynamicsStroke VolumeHemodynamic Monitoring

Stichwörter der Autor:innen

USCOMcardiac outputechocardiographyleft ventricular outflow tract velocitystroke volume

Autor:innen und Institutionen

  • Division of Pediatric Cardiology, Department of Pediatrics, Faculty of Medicine, Chiang Mai University, Chiang Mai, Thailand.
  • Department of Pediatrics, Faculty of Medicine, Chiang Mai University, Chiang Mai, Thailand.
  • Division of Pediatric Critical Care Medicine, Department of Pediatrics, Faculty of Medicine, Chiang Mai University, Chiang Mai, Thailand.

Zitieren

APA

Sethasathien, S., Arsasuwan, O., Jariyasakoolroj, T., & Silvilairat, S. (2026). Agreement Between Echocardiography and Ultrasonic Cardiac Output Monitoring (USCOM) for Hemodynamic Assessment in Pediatric Patients: A Bland–Altman Analysis. Echocardiography. https://doi.org/10.1111/echo.70645

Vancouver

Sethasathien S, Arsasuwan O, Jariyasakoolroj T, Silvilairat S. Agreement Between Echocardiography and Ultrasonic Cardiac Output Monitoring (USCOM) for Hemodynamic Assessment in Pediatric Patients: A Bland–Altman Analysis. Echocardiography. 2026. doi:10.1111/echo.70645

BibTeX
@article{Sethasathien2026,
  author = {Sethasathien, Saviga and Arsasuwan, Orana and Jariyasakoolroj, Theerapon and Silvilairat, Suchaya},
  title = {Agreement Between Echocardiography and Ultrasonic Cardiac Output Monitoring (USCOM) for Hemodynamic Assessment in Pediatric Patients: A Bland–Altman Analysis},
  journal = {Echocardiography},
  year = {2026},
  doi = {10.1111/echo.70645}
}