Effects of Electrode Thickness on Tricuspid Regurgitation and Radiation Exposure in Single‐Chamber Implantable Cardioverter Defibrillators

Fatih Gülçebi iD, Ahmet Özderya iD, İsmet Durmuş, Muhammet Raşit Sayın

Echocardiography · 2025

DOI: 10.1111/echo.70275 ↗

At a glance

Journal article in Echocardiography, 2025. Written by Fatih Gülçebi and 3 co-authors. Institution: Department of Cardiology, Trabzon Kanuni Training and Research Hospital, Trabzon, Turkey. Major MEDLINE topics: Tricuspid Valve Insufficiency, Defibrillators, Implantable, Echocardiography, Radiation Exposure. Cited 1 times so far (Crossref); 20 references.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

Currently, cardiac devices are widely used, and the effects of the electrodes of these devices on the tricuspid valve are being investigated. This study aimed to prospectively investigate, using two‐dimensional echocardiography, the impact of electrode thickness on the development of tricuspid regurgitation (TR) and operator radiation exposure during the implantation of single‐chamber implantable cardioverter‐defibrillators (VVI‐ICDs).

Methods

This prospective study included 61 patients treated at a single center between March 2023 and January 2024. Among these, 31 patients received 9F electrodes, and 30 patients received 7F electrodes. Initial echocardiographic evaluation was conducted 1 day before device implantation (T0). Follow‐up evaluations were performed 1 day (T1) and 1 month (T2) after implantation. Tricuspid regurgitation was assessed using two‐dimensional (2D) echocardiography and classified into grades 1–4 based on vena contracta and proximal isovelocity surface area measurements. A worsening of TR by one grade or more was deemed significant. Electrode implantation times and radiation doses were recorded during the procedures. At the 1‐month follow‐up, device interrogation was performed, and pacing percentages were documented.

Results

Baseline echocardiographic parameters were comparable between the two groups. No significant TR worsening was observed at T1 in either group ( p = 1.00). At T2, TR worsening was significantly more prevalent in the 9F group ( p = 0.033). However, the 7F group demonstrated significantly longer electrode implantation times ( p = 0.014) and higher total radiation exposure ( p = 0.011). No deterioration in right heart function was observed in either group during the 1‐month follow‐up.

Conclusions

The use of thinner electrodes in VVI‐ICD implantation may reduce electrode‐related TR worsening. However, thinner electrodes are more challenging to implant, resulting in prolonged procedure times and increased radiation exposure for the operator. Further large‐scale prospective studies are needed to confirm these findings.

Subjects

MeSH terms (MEDLINE)

Tricuspid Valve InsufficiencyDefibrillators, ImplantableEchocardiographyRadiation ExposureHumansMaleFemaleProspective StudiesMiddle AgedAgedTricuspid ValveElectrodesFollow-Up Studies

Author keywords

electrode thicknesssingle‐chamber implantable cardioverter‐defibrillatortricuspid regurgitation

Authors and institutions

  • Department of Cardiology, Trabzon Kanuni Training and Research Hospital, Trabzon, Turkey.
  • Department of Cardiology, University of Health Sciences Trabzon Ahi Evren Thoracic and Cardiovascular Surgery Training and Research Hospital, Trabzon, Turkey.

ORCID profiles: Fatih Gülçebi, Ahmet Özderya

Cite

APA

Gülçebi, F., Özderya, A., Durmuş, İ., & Sayın, M. R. (2025). Effects of Electrode Thickness on Tricuspid Regurgitation and Radiation Exposure in Single‐Chamber Implantable Cardioverter Defibrillators. Echocardiography. https://doi.org/10.1111/echo.70275

Vancouver

Gülçebi F, Özderya A, Durmuş İ, Sayın MR. Effects of Electrode Thickness on Tricuspid Regurgitation and Radiation Exposure in Single‐Chamber Implantable Cardioverter Defibrillators. Echocardiography. 2025. doi:10.1111/echo.70275

BibTeX
@article{Glebi2025,
  author = {Gülçebi, Fatih and Özderya, Ahmet and Durmuş, İsmet and Sayın, Muhammet Raşit},
  title = {Effects of Electrode Thickness on Tricuspid Regurgitation and Radiation Exposure in Single‐Chamber Implantable Cardioverter Defibrillators},
  journal = {Echocardiography},
  year = {2025},
  doi = {10.1111/echo.70275}
}
Start for free →