Feasibility of a Cardiac Magnetic Resonance Protocol for “off‐on” Cardiac Resynchronization Therapy Evaluation
Echocardiography · 2024
DOI: 10.1111/echo.70003 ↗At a glance
Studienprotokoll in Echocardiography, 2024. Written by Christian Basile and 11 co-authors. Institution: Department of Advanced Biomedical Sciences, Federico II University of Naples, Naples, Italy. Major MEDLINE topics: Cardiac Resynchronization Therapy, Feasibility Studies, Magnetic Resonance Imaging, Cine, Heart Failure. Cited 1 times so far (Crossref); 31 references.
Abstract (original language)
Background
Cardiac resynchronization therapy (CRT) is a standard treatment for patients with heart failure and electrical dyssynchrony. Cardiac magnetic resonance (CMR) is the gold standard for assessing left ventricular (LV) function. However, the feasibility of using CMR with active CRT is still uncertain.
Purpose
To assess the feasibility of a CRT “off‐on” protocol during CMR and measure the acute effects of CRT interruption on LV function.
Methods
Patients underwent CMR before (pre‐CRT) and 6 months after (post‐CRT) an MR‐conditional CRT defibrillator implantation. The post‐CRT scan included two complete sets of cine images, one with inactive (post‐CRT OFF ) and one with active CRT (post‐CRT ON ), maintaining a continuous connection between device and programmer.
Results
Out of 29 enrolled patients, 8 (28%) had complete and analyzable post‐CRT data. Unsuccessful procedures were attributed to connection problems between the CRT device and the programmer ( n = 10), poor image quality ( n = 7), and lack of patient cooperation ( n = 4). LV ejection fraction significantly increased between pre‐CRT scan (28.1%) and both post‐CRT OFF (37.9%; p = 0.046) and post‐CRT ON CMR (35.0%; p = 0.037), with a nonstatistically significant trend toward decreased LV volumes. No adverse events or significant changes in device electrical parameters (including battery level) were detected during the post‐CMR scan period.
Conclusion
A CRT “off‐on” protocol during CMR studies can be safely executed in patients with an MR‐conditional CRT defibrillator. However, technical improvements are needed to facilitate high‐quality scans during active CRT. Favorable changes in LV function induced by CRT remodeling were not acutely reversed with the interruption of electrical therapy.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Department of Advanced Biomedical Sciences, Federico II University of Naples, Naples, Italy.
- Division of Cardiology, Department of Medicine, Karolinska Institutet, Stockholm, Sweden.
- Clinica Villa dei Fiori, Advanced Cardiovascular Imaging Unit, Acerra, Naples, Italy.
- Department of Medical, Motor and Wellness Sciences, Parthenope University of Naples, Naples, Italy.
- Clinical Unit, Biotronik Italia S.p.a., Cologno Monzese, Italy.
ORCID profiles: Christian Basile, Costantino Mancusi
Cite
APA
Basile, C., Scatteia, A., Giacopelli, D., Gallo, P., Pezzullo, S., Mancusi, C., Pascale, C. E., Gargiulo, P., Marzano, F., Perrone-Filardi, P., Paolillo, S., & Dellegrottaglie, S. (2024). Feasibility of a Cardiac Magnetic Resonance Protocol for “off‐on” Cardiac Resynchronization Therapy Evaluation. Echocardiography. https://doi.org/10.1111/echo.70003
Vancouver
Basile C, Scatteia A, Giacopelli D, Gallo P, Pezzullo S, Mancusi C, et al. Feasibility of a Cardiac Magnetic Resonance Protocol for “off‐on” Cardiac Resynchronization Therapy Evaluation. Echocardiography. 2024. doi:10.1111/echo.70003
BibTeX
@article{Basile2024,
author = {Basile, Christian and Scatteia, Alessandra and Giacopelli, Daniele and Gallo, Paolo and Pezzullo, Salvatore and Mancusi, Costantino and Pascale, Carmine E and Gargiulo, Paola and Marzano, Federica and Perrone-Filardi, Pasquale and Paolillo, Stefania and Dellegrottaglie, Santo},
title = {Feasibility of a Cardiac Magnetic Resonance Protocol for “off‐on” Cardiac Resynchronization Therapy Evaluation},
journal = {Echocardiography},
year = {2024},
doi = {10.1111/echo.70003}
}
EUVIMED