Transthoracic echocardiographic Doppler metrics in evaluating bioprosthetic tricuspid valve dysfunction
Echocardiography · 2024
DOI: 10.1111/echo.15835 ↗At a glance
Journal article in Echocardiography, 2024. Written by Xin Zhang and 11 co-authors. Institution: Department of Ultrasound Medicine, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, China. Major MEDLINE topics: Tricuspid Valve, Bioprosthesis, Echocardiography, Doppler, Heart Valve Prosthesis. Cited 1 times so far (Crossref); 24 references. Funded by: the National Natural Science Foundation of China, the Nature Science Foundation of Hubei Province.
Abstract (original language)
Purpose
There is currently limited information on the utility of transthoracic echocardiography (TTE)‐derived Doppler parameters for assessing bioprosthetic tricuspid valve (BTV) dysfunction. Our study aimed to establish the precision and appropriate reference ranges for routinely collected transthoracic Doppler parameters in the assessment of BTV dysfunction.
Methods
We retrospectively evaluated 100 BTV patients who underwent TTE. Based on redo surgical confirmation or more than 2 repeat TTE or transesophageal echocardiography (TEE) examinations, patients were allocated to normal ( n = 61), regurgitant ( n = 24), or stenotic ( n = 15) BTV group. Univariate and multivariate binary logistic regression were performed to identify TTE Doppler parameters that detected BTV dysfunction.
Results
The VTI ratio (VTI TV /VTI LVOT ) was the most accurate Doppler parameter for detecting BTV dysfunction, with a ratio of >2.8 showing 84.6% sensitivity and 90.2% specificity. VTI ratio > 3.2, mean gradient (MG TV ) > 6.2 mmHg and pressure half‐time > 218 ms detected significant BTV stenosis, with sensitivities of 100%, 93.3% and 93.3% and specificities of 82.4%, 75.3% and 87.1%, respectively. After multivariate analysis, the VTI ratio > 2.8 (OR = 9.00, 95% CI = 2.13–41.61, p = .003) and MG TV > 5.1 mmHg (OR = 6.50, 95% CI = 1.69–27.78, p = .008) were the independent associations of BTV dysfunction. With these cutoff values, 75.0%–92.2% of normal and 62.5%–96.0% of dysfunctional BTV were identified.
Conclusions
Doppler parameters from TTE can accurately identify BTV dysfunction, particularly with VTI ratio > 2.8 and MG TV > 5.1 mmHg, to assess the need for additional testing with TEE.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Department of Ultrasound Medicine, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, China.
- Hubei Province Clinical Research Center for Medical Imaging, Wuhan, China.
- Hubei Province Key Laboratory of Molecular Imaging, Wuhan, China.
ORCID profiles: Jia Xu, Jing Wang, Li Zhang, Qing Lv, Mingxing Xie, Wenqian Wu
Funding
- the National Natural Science Foundation of China
- the Nature Science Foundation of Hubei Province
Cite
APA
Zhang, X., Peng, L., Fang, L., Xu, J., Wang, J., Sun, W., Gao, T., Li, Y., Zhang, L., Lv, Q., Xie, M., & Wu, W. (2024). Transthoracic echocardiographic Doppler metrics in evaluating bioprosthetic tricuspid valve dysfunction. Echocardiography. https://doi.org/10.1111/echo.15835
Vancouver
Zhang X, Peng L, Fang L, Xu J, Wang J, Sun W, et al. Transthoracic echocardiographic Doppler metrics in evaluating bioprosthetic tricuspid valve dysfunction. Echocardiography. 2024. doi:10.1111/echo.15835
BibTeX
@article{Zhang2024,
author = {Zhang, Xin and Peng, Lingli and Fang, Lingyun and Xu, Jia and Wang, Jing and Sun, Wei and Gao, Tang and Li, Yuman and Zhang, Li and Lv, Qing and Xie, Mingxing and Wu, Wenqian},
title = {Transthoracic echocardiographic Doppler metrics in evaluating bioprosthetic tricuspid valve dysfunction},
journal = {Echocardiography},
year = {2024},
doi = {10.1111/echo.15835}
}
EUVIMED