Performance of exercise transcutaneous oximetry versus imaging at the buttock, thigh and calf level for the diagnosis of peripheral artery disease
Clinical Physiology and Functional Imaging · 2026-05 · 46 (3)
DOI: 10.1111/cpf.70068 ↗At a glance
Vergleichsstudie in Clinical Physiology and Functional Imaging, published online on 2026-05-18. Written by Mariève Houle and 5 co-authors. Institution: Université d'Angers, Inserm, CNRS, MITOVASC, Equipe CarMe, SFR ICAT, Angers, France. Major MEDLINE topics: Blood Gas Monitoring, Transcutaneous, Peripheral Arterial Disease, Thigh, Intermittent Claudication. 34 references. The full text is freely available (open access, CC BY 4.0).
Abstract (original language)
Background
Exercise transcutaneous oximetry (Ex‐TcPO 2 ), through the minimal value of the “decrease from rest of oxygen pressure” (DROPmin), identifies ischemia during exercise in patients with PAD. However, its diagnostic performance at the thigh level compared to imaging modalities, has not been evaluated. The objective of this retrospective study was to perform an external validation of Ex‐TcPO₂ results at the buttock and calf, a first validation of Ex‐TcPO₂ at the thigh level and to determine the respective diagnostic performance of the DROPmin and DROPend (the DROP at the time of exercise cessation).
Methods
We studied 192 patients complaining of intermittent claudication. We retrieved DROPmin and DROPend values from Ex‐TcPO 2 testing and the presence/absence of an ipsilateral significant stenosis ( ≥ 70%) at imaging.
Results
The performance of Ex‐TcPO₂ using DROPmin values at the buttock, thigh and calf levels were respectively fair (AUC = 0.70; 95% CI: 0.64–0.76), fair (AUC = 0.77; 95% CI: 0.72–0.83) and excellent (AUC = 0.93; 95% CI: 0.90–0.96). When using DROPend, the performance of Ex‐TcPO 2 was no more significant. Discussion/Conclusion This study showed that Ex‐TcPO₂ has a fair to excellent accuracy for detecting the presence of at least one artery stenosis ≥70% in the lower limb according to buttock, thigh and calf area in highly selected patients. This study also highlights the importance of using the minimal value of DROPmin rather than the DROPend.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Université d'Angers, Inserm, CNRS, MITOVASC, Equipe CarMe, SFR ICAT, Angers, France.
- Service of Sports Medicine, Angers University Hospital, Angers, France.
- Service of Vascular Medicine, Angers University Hospital, Angers, France.
- School of Electronics (ESEO), Angers, France.
- UMR CNRS, Le Mans, France.
ORCID profiles: Mariève Houle, Pierre Abraham
Cite
APA
Houle, M., Lecoq, S., Feuilloy, M., Hersant, J., Henni, S., & Abraham, P. (2026). Performance of exercise transcutaneous oximetry versus imaging at the buttock, thigh and calf level for the diagnosis of peripheral artery disease. Clinical Physiology and Functional Imaging, 46(3). https://doi.org/10.1111/cpf.70068
Vancouver
Houle M, Lecoq S, Feuilloy M, Hersant J, Henni S, Abraham P. Performance of exercise transcutaneous oximetry versus imaging at the buttock, thigh and calf level for the diagnosis of peripheral artery disease. Clinical Physiology and Functional Imaging. 2026;46(3). doi:10.1111/cpf.70068
BibTeX
@article{Houle2026,
author = {Houle, Mariève and Lecoq, Simon and Feuilloy, Mathieu and Hersant, Jeanne and Henni, Samir and Abraham, Pierre},
title = {Performance of exercise transcutaneous oximetry versus imaging at the buttock, thigh and calf level for the diagnosis of peripheral artery disease},
journal = {Clinical Physiology and Functional Imaging},
year = {2026},
volume = {46},
number = {3},
doi = {10.1111/cpf.70068}
}
EUVIMED