Reproducibility of diastolic function parameters in repeated ECG‐gated SPECT myocardial perfusion imaging and echocardiography
Clinical Physiology and Functional Imaging · 2025
DOI: 10.1111/cpf.70032 ↗At a glance
Vergleichsstudie in Clinical Physiology and Functional Imaging, 2025. Written by Aaro Krohns and 3 co-authors. Institution: University of Eastern Finland, Kuopio, Finland. Major MEDLINE topics: Myocardial Perfusion Imaging, Cardiac-Gated Single-Photon Emission Computer-Assisted Tomography, Ventricular Function, Left, Electrocardiography. 30 references. The full text is freely available (open access, CC BY-NC 4.0). Funded by: This study was financially supported by the Finnish Foundation for Cardiovascular Research (10.11.2020 for T.P.L.). S.S. has achieved funding from the Research….
Abstract (original language)
Objective
To study the reproducibility of diastolic function parameters from myocardial perfusion imaging (MPI) using ECG‐gated single photon emission computed tomography (SPECT), and echocardiography in repeated imaging.
Methods
We studied the reproducibility of MPI diastolic function parameters peak‐filling rate (PFR) and time to peak filling (TTPF) as well as echocardiographic parameters E/A, E/e', and E‐wave deceleration time (DT). The study population consisted of 21 patients who underwent routine MPI with repeated rest acquisition and echocardiography. In a test–retest setting, appropriate diastolic measures were successfully obtained using SPECT in 20 patients, with E/A measured in 18, E/e' in 17, and DT in 16 patients.
Results
We found that PFR was well reproducible with the intraclass correlation coefficient (ICC) 0.887, and coefficient of variation (CV%) 10.5. However, TTPF was poorly reproducible (ICC 0.012, and CV% 17.5). E/A was highly reproducible (ICC 0.963, and CV% 12.5). Decent reproducibility was observed for E/e' (ICC 0.809, and CV% 18.6) and DT (ICC 0.833, and CV% 12.0). We further studied the correlation between these parameters. PFR (EDV/s) had negative correlation with DT (r = −0.538, p = 0.039) and E/A had positive with E/é ( r = 0.689, p = 0.002). No other significant correlations were observed ( p > 0.05 for all). We also examined how parameters classified patients as normal or abnormal regarding the diastolic function. E/A and E/e´, as well as E/A and PFR showed similar classifications in 88% of cases, with a Kappa value of 0.433, p = 0.074 for both.
Conclusions
PFR obtained from repeated SPECT studies, along with the E/A ratio, is highly reproducible.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- University of Eastern Finland, Kuopio, Finland.
- Diagnostic Imaging Center, Kuopio University Hospital, Kuopio, Finland.
- Institute of Clinical Medicine, University of Eastern Finland, Kuopio, Finland.
ORCID profiles: Saara Sillanmäki
Funding
- This study was financially supported by the Finnish Foundation for Cardiovascular Research (10.11.2020 for T.P.L.). S.S. has achieved funding from the Research…
Cite
APA
Krohns, A., Laitinen, T. P., Laitinen, T. M., & Sillanmäki, S. (2025). Reproducibility of diastolic function parameters in repeated ECG‐gated SPECT myocardial perfusion imaging and echocardiography. Clinical Physiology and Functional Imaging. https://doi.org/10.1111/cpf.70032
Vancouver
Krohns A, Laitinen TP, Laitinen TM, Sillanmäki S. Reproducibility of diastolic function parameters in repeated ECG‐gated SPECT myocardial perfusion imaging and echocardiography. Clinical Physiology and Functional Imaging. 2025. doi:10.1111/cpf.70032
BibTeX
@article{Krohns2025,
author = {Krohns, Aaro and Laitinen, Tomi P and Laitinen, Tiina M and Sillanmäki, Saara},
title = {Reproducibility of diastolic function parameters in repeated ECG‐gated SPECT myocardial perfusion imaging and echocardiography},
journal = {Clinical Physiology and Functional Imaging},
year = {2025},
doi = {10.1111/cpf.70032}
}
EUVIMED