Implant‐Supported Cantilever Fixed Dental Prosthesis in the Anterior Region: Effect of Implant Type and Aging In Vitro
Clinical Oral Implants Research · 2025
DOI: 10.1111/clr.14388 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinical Oral Implants Research, 2025. Verfasst von Nadja Rohr und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Reconstructive Dentistry, University Center for Dental Medicine Basel UZB, University of Basel, Basel, Switzerland. Hauptthemen laut MEDLINE: Dental Prosthesis, Implant-Supported, Dental Implants, Denture, Partial, Fixed. Die Arbeit nennt 31 Literaturangaben. Gefördert von: ITI Foundation.
Abstract (Originalsprache)
Objectives
To investigate if the fracture load of implant‐supported cantilever fixed dental prostheses (ICFDPs) in the anterior region is affected by the implant type and the aging protocol. Materials and Methods Sixty ICFDPs were prepared using multilayer monolithic zirconia for restoring bone‐level (BL) and tissue‐level (TL) titanium‐zirconium implants. Fracture load was measured at baseline (no aging) or after aging in a chewing simulator loading the implant crown or the cantilever in a 30° angle ( n = 10 per group). A two‐way ANOVA was applied ( α = 0.05) for the effects of implant type and aging protocol.
Results
Implant type ( p < 0.001) and aging ( p < 0.001) had a significant effect on fracture load values. Baseline specimens of both BL (665 ± 62 N) and TL (554 ± 23 N) had significantly higher mean fracture load values than those of aged specimens(BL implant crown 545 ± 46 N, BL cantilever 563 ± 45 N, TL implant crown 455 ± 32 N, TL cantilever 476 ± 24 N) ( p < 0.001). The loading position during aging did not affect fracture load values for BL ( p = 0.980) nor TL ( p = 0.749). BL implants failed by cement fracture and abutment deformation, while for TL the transmucosal part of the implant deformed.
Conclusions
The fracture load values of anterior ICFDPs decrease with aging in a chewing simulator. BL implants are preferable over TL implants for anterior ICFDPs as higher fracture load values were achieved, and no implant deformation occurred.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Reconstructive Dentistry, University Center for Dental Medicine Basel UZB, University of Basel, Basel, Switzerland.
- Biomaterials and Technology, Department of Research, University Center for Dental Medicine Basel UZB, University of Basel, Basel, Switzerland.
ORCID-Profile: Nadja Rohr, Yotaro Iwauchi, Nicola U Zitzmann
Förderung
- ITI Foundation
Zitieren
APA
Rohr, N., Iwauchi, Y., Bernauer, R., Heuzeroth, R., & Zitzmann, N. U. (2025). Implant‐Supported Cantilever Fixed Dental Prosthesis in the Anterior Region: Effect of Implant Type and Aging In Vitro. Clinical Oral Implants Research. https://doi.org/10.1111/clr.14388
Vancouver
Rohr N, Iwauchi Y, Bernauer R, Heuzeroth R, Zitzmann NU. Implant‐Supported Cantilever Fixed Dental Prosthesis in the Anterior Region: Effect of Implant Type and Aging In Vitro. Clinical Oral Implants Research. 2025. doi:10.1111/clr.14388
BibTeX
@article{Rohr2025,
author = {Rohr, Nadja and Iwauchi, Yotaro and Bernauer, Raffael and Heuzeroth, Raphael and Zitzmann, Nicola U},
title = {Implant‐Supported Cantilever Fixed Dental Prosthesis in the Anterior Region: Effect of Implant Type and Aging In Vitro},
journal = {Clinical Oral Implants Research},
year = {2025},
doi = {10.1111/clr.14388}
}
EUVIMED