Influence of Metal Artifact Reduction Tool of Two Cone Beam CT on the Detection of Bone Graft Loss Around Titanium and Zirconium Implants—An Ex Vivo Diagnostic Accuracy Study
Clinical Oral Implants Research · 2025
DOI: 10.1111/clr.14381 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinical Oral Implants Research, 2025. Verfasst von Michelle Chang und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Dental Materials and Prosthesis, Ribeirão Preto School of Dentistry, University of São Paulo (USP), Ribeirão Preto, São Paulo, Brazil. Hauptthemen laut MEDLINE: Titanium, Cone-Beam Computed Tomography, Zirconium, Artifacts. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 59 Literaturangaben. Gefördert von: Coordenação de Aperfeiçoamento de Pessoal de Nível Superior, Fundação de Amparo à Pesquisa do Estado de São Paulo.
Abstract (Originalsprache)
Objective
The objective of this study is to investigate the influence of metal artifact reduction (MAR) on two cone beam computed tomography (CBCT) units in the evaluation of bone graft loss adjacent to titanium (Ti) and zirconium (Zr) implants. Material and Methods Twelve Ti and twelve Zr implants were placed in the posterior region of dry human mandibles. Bone graft was applied to the level of the cover screw. Bone graft loss was simulated in half of the sample (6 Ti and 6 Zr) by removing the graft material up to the third implant thread on the buccal surface. CBCT images were acquired on two units, varying the application of MAR (OP300—off and on; Eagle 3D—standard, intermediate, and extreme). The images were assessed by five evaluators that scored the presence of graft loss according to a 5‐point scale. The diagnostic values were calculated and compared by non‐parametric tests with a significance level of 5%.
Results
Higher diagnostic values were achieved with MAR activated in the OP300 unit, for Ti and Zr (p < 0.05) . On the Eagle 3D unit, MAR in extreme mode resulted in lower diagnostic values for both types of implants (p < 0.05) . The diagnostic values of Ti implants were higher than Zr implants (p < 0.05) .
Conclusion
The application of MAR influences the diagnosis of bone graft loss adjacent to Ti and Zr dental implants. However, the extreme mode of MAR in the Eagle 3D unit can impair the diagnostic task in both types of implants and should be avoided.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Dental Materials and Prosthesis, Ribeirão Preto School of Dentistry, University of São Paulo (USP), Ribeirão Preto, São Paulo, Brazil.
- Department of Diagnosis & Oral Health, University of Louisville School of Dentistry, Louisville, Kentucky, USA.
- Department of Stomatology, Public Health, and Forensic Dentistry, Ribeirão Preto School of Dentistry, University of São Paulo (USP), Ribeirão Preto, São Paulo, Brazil.
ORCID-Profile: Michelle Chang
Förderung
- Coordenação de Aperfeiçoamento de Pessoal de Nível Superior
- Fundação de Amparo à Pesquisa do Estado de São Paulo
Zitieren
APA
Chang, M., Felizardo, H. M. A., Oliveira-Santos, C., & Gaêta-Araujo, H. (2025). Influence of Metal Artifact Reduction Tool of Two Cone Beam CT on the Detection of Bone Graft Loss Around Titanium and Zirconium Implants—An Ex Vivo Diagnostic Accuracy Study. Clinical Oral Implants Research. https://doi.org/10.1111/clr.14381
Vancouver
Chang M, Felizardo HMA, Oliveira-Santos C, Gaêta-Araujo H. Influence of Metal Artifact Reduction Tool of Two Cone Beam CT on the Detection of Bone Graft Loss Around Titanium and Zirconium Implants—An Ex Vivo Diagnostic Accuracy Study. Clinical Oral Implants Research. 2025. doi:10.1111/clr.14381
BibTeX
@article{Chang2025,
author = {Chang, Michelle and Felizardo, Henrique Mateus Alves and Oliveira-Santos, Christiano and Gaêta-Araujo, Hugo},
title = {Influence of Metal Artifact Reduction Tool of Two Cone Beam CT on the Detection of Bone Graft Loss Around Titanium and Zirconium Implants—An Ex Vivo Diagnostic Accuracy Study},
journal = {Clinical Oral Implants Research},
year = {2025},
doi = {10.1111/clr.14381}
}
EUVIMED