Single‐piece zirconia versus single‐piece titanium, narrow‐diameter dental implants in the anterior maxilla: 5‐year post‐loading results of a randomized clinical trial

Paula Andrea Ruiz Henao iD, Gabriel Leonardo Magrin, Leticia Caneiro-Queija, Cesar Augusto Magalhães Benfatti iD, Yago Leira iD, Antonio Liñares-González iD, Juan Blanco-Carrión iD

Clinical Oral Implants Research · 2024

DOI: 10.1111/clr.14319 ↗

Kurz gefasst

Randomisierte kontrollierte Studie in der Zeitschrift Clinical Oral Implants Research, 2024. Verfasst von Paula Andrea Ruiz Henao und 6 weiteren Autor:innen. Beteiligte Einrichtung: Unit of Periodontology, Department of Surgery and Medical-Surgical Specialties, Faculty of Medicine and Dentistry, University of Santiago de Composte…. Hauptthemen laut MEDLINE: Zirconium, Titanium, Maxilla, Esthetics, Dental. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 44 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Institut Straumann AG, Institute of Health Carlos III, Straumann Group.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

The aim of this study was to evaluate esthetic parameters in the anterior maxillary region by comparing single‐piece zirconia versus titanium narrow‐diameter implants. Additionally, clinical, radiological and patient‐reported outcome measures (PROMs) were analyzed. Materials and Methods Thirty implants (tissue level implant) were placed in 30 patients in the maxillary esthetic sector. Depending on randomization, a zirconia (test) or titanium implant (control) was placed. Esthetic, clinical, and radiological parameters, including the implant crown esthetic index (ICAI), pink esthetic score (PES), probing pocket depth, bleeding on probing, plaque index, and marginal bone levels, were evaluated at 12, 36 and 60 months after loading.

Results

Sixty months after crown placement, no significant differences were found between groups. The ICAI values were 5.25 ± 4.21 and 4.50 ± 2.98 for the test and control groups, respectively. The corresponding PES values were 7.44 ± 1.93 and 7.43 ± 1.74 for the test and control groups, respectively. There were no significant intergroup differences for the rest of the parameters evaluated.

Conclusion

It can be suggested that monotype zirconia implants may serve as a potential alternative to titanium implants in selected clinical scenarios. While the results demonstrated comparable esthetic, clinical, and radiological aspects for zirconia implants as compared to titanium implants after a 5‐year follow‐up period, further research with larger sample sizes and longer‐term follow‐up is recommended.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ZirconiumTitaniumMaxillaEsthetics, DentalDental Prosthesis DesignHumansFemaleMaleMiddle AgedAdultDental ImplantsDental Implantation, Endosseous

Stichwörter der Autor:innen

dental implantsestheticslongitudinal studiesrandomized controlled trialsingle‐tooth dental implantszirconia dental implants

Substanzen

Zirconiumzirconium oxideTitaniumDental Implants

Autor:innen und Institutionen

  • Unit of Periodontology, Department of Surgery and Medical-Surgical Specialties, Faculty of Medicine and Dentistry, University of Santiago de Compostela, Santiago de Compostela, Spain.
  • Center for Education and Research on Dental Implants (CEPID), Department of Dentistry, Federal University of Santa Catarina (UFSC), Florianópolis, Brazil.

ORCID-Profile: Paula Andrea Ruiz Henao, Cesar Augusto Magalhães Benfatti, Yago Leira, Antonio Liñares-González, Juan Blanco-Carrión

Förderung

  • Institut Straumann AG
  • Institute of Health Carlos III
  • Straumann Group
  • Instituto de Salud Carlos III

Zitieren

APA

Ruiz Henao, P. A., Magrin, G. L., Caneiro-Queija, L., Benfatti, C. A. M., Leira, Y., Liñares-González, A., & Blanco-Carrión, J. (2024). Single‐piece zirconia versus single‐piece titanium, narrow‐diameter dental implants in the anterior maxilla: 5‐year post‐loading results of a randomized clinical trial. Clinical Oral Implants Research. https://doi.org/10.1111/clr.14319

Vancouver

Ruiz Henao PA, Magrin GL, Caneiro-Queija L, Benfatti CAM, Leira Y, Liñares-González A, et al. Single‐piece zirconia versus single‐piece titanium, narrow‐diameter dental implants in the anterior maxilla: 5‐year post‐loading results of a randomized clinical trial. Clinical Oral Implants Research. 2024. doi:10.1111/clr.14319

BibTeX
@article{RuizHenao2024,
  author = {Ruiz Henao, Paula Andrea and Magrin, Gabriel Leonardo and Caneiro-Queija, Leticia and Benfatti, Cesar Augusto Magalhães and Leira, Yago and Liñares-González, Antonio and Blanco-Carrión, Juan},
  title = {Single‐piece zirconia versus single‐piece titanium, narrow‐diameter dental implants in the anterior maxilla: 5‐year post‐loading results of a randomized clinical trial},
  journal = {Clinical Oral Implants Research},
  year = {2024},
  doi = {10.1111/clr.14319}
}
Kostenlos starten →