Influence of aging of PEEK attachment inserts on the pull‐off force of implant‐retained overdentures – A laboratory study

Andreas Koenig, Laura Rotenburg, Florian Fuchs, Stephan Sander, Bernd Lethaus, Sebastian Hahnel

Clinical Oral Implants Research · 2023

DOI: 10.1111/clr.14180 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Clinical Oral Implants Research, 2023. Verfasst von Andreas Koenig und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Prosthodontics and Material Sciences, Leipzig University, Leipzig, Germany. Hauptthemen laut MEDLINE: Denture Retention, Denture, Overlay. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: International Team for Implantology.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

The aim of the current study is to investigate the influence of mechanical stress as well as cleaning agents on the performance of various polyether ether ketone (PEEK) inserts for implant‐retained overdentures (IOD). Materials and Methods Three different standard PEEK inserts were subjected to rapid artificial aging through storage in chemical denture cleaning agents (acetic acid, sodium hypochlorite, or sparkling denture cleaner) as well as demineralized water. The pre‐aged PEEK inserts were then placed in unilateral IOD and subjected to 200,000 chewing loads (5 kg ~ 50 N), with 5000 thermal cycles (5/55°C), and 1100 removal/insertion cycles (vertical movement 2 mm).

Results

For all the PEEK inserts, the retention forces decreased significantly with an increasing number of mechanical load cycles and after exposure to all the cleaning agents. PEEK inserts aged by exposure to chemical cleaning agents showed a significantly higher decline in retention force than the inserts stored in water. Confocal laser scanning microscopy indicated that the decline in retention force might be caused by wear on the internal insert surface in contact with the patrix.

Conclusions

Within the limitations of this study, it can be concluded that the application of chemical cleaning agents accelerates the decline in the retention forces of PEEK retentive inserts in IODs.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Denture RetentionDenture, OverlayDental Prosthesis, Implant-SupportedDental Stress AnalysisBenzophenonesPolyethylene GlycolsWaterPolymers

Stichwörter der Autor:innen

Novaloc®chemical cleansingchewing simulationpolyaryl ether ketonepolyether ether ketoneretention force

Substanzen

BenzophenonesPolyethylene GlycolsWaterpolyetheretherketonePolymers

Autor:innen und Institutionen

  • Department of Prosthodontics and Material Sciences, Leipzig University, Leipzig, Germany.
  • Department of Cranio Maxillofacial Surgery, Leipzig, Germany.
  • Department of Prosthetic Dentistry, UKR University Hospital Regensburg, Regensburg, Germany.

Förderung

  • International Team for Implantology

Zitieren

APA

Koenig, A., Rotenburg, L., Fuchs, F., Sander, S., Lethaus, B., & Hahnel, S. (2023). Influence of aging of PEEK attachment inserts on the pull‐off force of implant‐retained overdentures – A laboratory study. Clinical Oral Implants Research. https://doi.org/10.1111/clr.14180

Vancouver

Koenig A, Rotenburg L, Fuchs F, Sander S, Lethaus B, Hahnel S. Influence of aging of PEEK attachment inserts on the pull‐off force of implant‐retained overdentures – A laboratory study. Clinical Oral Implants Research. 2023. doi:10.1111/clr.14180

BibTeX
@article{Koenig2023,
  author = {Koenig, Andreas and Rotenburg, Laura and Fuchs, Florian and Sander, Stephan and Lethaus, Bernd and Hahnel, Sebastian},
  title = {Influence of aging of PEEK attachment inserts on the pull‐off force of implant‐retained overdentures – A laboratory study},
  journal = {Clinical Oral Implants Research},
  year = {2023},
  doi = {10.1111/clr.14180}
}
Kostenlos starten →