High‐Resolution Microperimetry for Detecting Glaucomatous Damage: A Prospective Evaluation of Performance
Clinical & Experimental Ophthalmology · 2025
DOI: 10.1111/ceo.14528 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Clinical & Experimental Ophthalmology, 2025. Verfasst von Marc G Sarossy und 9 weiteren Autor:innen. Beteiligte Einrichtung: Centre for Eye Research Australia, Royal Victorian Eye and Ear Hospital, East Melbourne, Victoria, Australia. Hauptthemen laut MEDLINE: Visual Field Tests, Visual Fields, Optic Nerve Diseases, Optic Disk. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 51 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: National Health & Medical Research Council of Australia, BrightFocus Foundation, F. Hoffmann-La Roche Ltd.
Abstract (Originalsprache)
Background
To compare the performance of high‐resolution microperimetry testing against standard automated perimetry (SAP) for detecting glaucomatous damage seen on optical coherence tomography (OCT) scans.
Methods
250 eyes from 200 individuals underwent high‐resolution microperimetry testing of a hemifield using a stimulus pattern optimised to sample typical arcuate patterns of visual field loss in glaucoma. SAP was performed using a 24–2 stimulus pattern. The presence of glaucomatous damage in each hemifield was subsequently independently assessed by two graders based on a circumpapillary OCT circle scan, 6×6 mm optic disc‐centred OCT volume scan, and 15×15 mm widefield OCT volume scan.
Results
The hemifield‐based mean total deviation (MTD) and pointwise sensitivity standard deviation (PSD) for high‐resolution microperimetry and SAP were both independently associated with the presence of glaucomatous damage in multivariable logistic regression analyses ( p ≤ 0.046). Prediction models developed using these two parameters showed significantly higher performance with high‐resolution microperimetry compared to SAP (partial area under the receiver operating characteristic curve [pAUC] = 0.86 and 0.75 respectively; p = 0.007). The performance of microperimetry decreased with decreasing resolution when the MTD and PSD values were derived from every second, third, and sixth test locations (pAUC = 0.85 [ p = 0.16], 0.85 [ p = 0.044] and 0.84 [ p < 0.001] respectively), but they were all significantly better than SAP ( p ≤ 0.028 for all).
Conclusions
High‐resolution microperimetry enabled improved performance for detecting glaucomatous damage compared to SAP, highlighting the potential value of fundus‐tracking and higher‐resolution sampling for the detection of glaucomatous visual field loss.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Centre for Eye Research Australia, Royal Victorian Eye and Ear Hospital, East Melbourne, Victoria, Australia.
- Ophthalmology, Department of Surgery, The University of Melbourne, Melbourne, Victoria, Australia.
ORCID-Profile: Zhichao Wu
Förderung
- National Health & Medical Research Council of Australia
- BrightFocus Foundation
- F. Hoffmann-La Roche Ltd
- Roche
Zitieren
APA
Sarossy, M. G., Rezk, S., Li, D., Hadoux, X., Jannaud, M., Coote, M. A., Ang, G. S., van Wijngaarden, P., Martin, K. R., & Wu, Z. (2025). High‐Resolution Microperimetry for Detecting Glaucomatous Damage: A Prospective Evaluation of Performance. Clinical & Experimental Ophthalmology. https://doi.org/10.1111/ceo.14528
Vancouver
Sarossy MG, Rezk S, Li D, Hadoux X, Jannaud M, Coote MA, et al. High‐Resolution Microperimetry for Detecting Glaucomatous Damage: A Prospective Evaluation of Performance. Clinical & Experimental Ophthalmology. 2025. doi:10.1111/ceo.14528
BibTeX
@article{Sarossy2025,
author = {Sarossy, Marc G and Rezk, Sandy and Li, Dongzhe and Hadoux, Xavier and Jannaud, Maxime and Coote, Michael A and Ang, Ghee Soon and van Wijngaarden, Peter and Martin, Keith R and Wu, Zhichao},
title = {High‐Resolution Microperimetry for Detecting Glaucomatous Damage: A Prospective Evaluation of Performance},
journal = {Clinical & Experimental Ophthalmology},
year = {2025},
doi = {10.1111/ceo.14528}
}
EUVIMED