In vitro comparison of stone retropulsion between the new holmium:yttrium aluminium garnet laser generator with a very low peak power pulse modulation mode (Magneto technology) and the thulium fibre laser
BJU International · 2026-02 · 137 (2) · 306-311
DOI: 10.1111/bju.70083 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift BJU International, online erschienen am 19.11.2025. Verfasst von Andrea Folcia und 12 weiteren Autor:innen. Beteiligte Einrichtung: Division of Experimental Oncology/Unit of Urology, Urological Research Institute, IRCCS Ospedale San Raffaele, Milan, Italy. Hauptthemen laut MEDLINE: Lasers, Solid-State, Thulium, Lithotripsy, Laser, Urinary Calculi. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 23 Literaturangaben.
Abstract (Originalsprache)
Objectives
To compare the stone retropulsion of an artificial model of urinary stones in vitro using the new holmium:yttrium aluminium garnet (Ho:YAG) laser generator with a very low peak power pulse modulation mode (Quanta Magneto Technology™; Quanta System, Samarate, Italy) and thulium fibre laser (TFL). Materials and Methods The following laser energy settings were used: 0.2 J–50 Hz, 0.4 J–25 Hz, 0.8 J–12 Hz, 1.5 J–6 Hz, and 2 J–5 Hz. A 5 × 5 × 5 mm cubic artificial stone was placed on two perpendicularly aligned metal rulers to create a V‐shaped cross‐sectional rail submerged in a tank filled with 0.9% saline solution, with the laser fibre in contact with the stone. The laser was activated for 5 s. Stone retropulsion was captured by a camera and assessed using video tracking software. Five repetitions were performed for each setting. Descriptive statistics investigated differences between the two lasers in stone retropulsion distance and speed at selected time points (0.35, 1, 3, and 5 s) and over the first second following laser activation.
Results
Comparable stone retropulsion distance and speed were observed between ‘Magneto’ and TFL according to tested high‐energy settings. Magneto exhibited lower mean stone displacement at selected time points (all P < 0.02) and lower mean retropulsion speed over the first second following laser activation (all P < 0.01), when compared to TFL at low‐energy settings (0.2 and 0.4 J).
Conclusions
Although Ho:YAG lasers are generally associated with greater stone retropulsion compared to TFL, Magneto exhibited a retropulsion performance similar to TFL. Further research is needed to clarify the retropulsion dynamics observed with Magneto at low‐energy settings.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Division of Experimental Oncology/Unit of Urology, Urological Research Institute, IRCCS Ospedale San Raffaele, Milan, Italy.
- Universita Vita-Salute San Raffaele University, Milan, Italy.
- Ospedale Sant'Anna Hospital, Turin, Italy.
- EAU Endourology, European Association of Urology (EAU), Arnhem, The Netherlands.
- Department of Urology, IRCCS Ospedale Galeazzi-Sant'Ambrogio, Milan, Italy.
- Istituto Clinico Beato Matteo, Vigevano, Italy.
ORCID-Profile: Andrea Folcia, Davide Perri, Luca Villa, Daniele Robesti, Giorgio Bozzini, Ioannis Kartalas Goumas, Francesco Montorsi, Andrea Salonia, Eugenio Ventimiglia
Zitieren
APA
Folcia, A., Perri, D., Villa, L., Albanese, N., Robesti, D., Corsini, C., Bozzini, G., Kartalas Goumas, I., Hota, T., Montorsi, F., Salonia, A., Traxer, O., & Ventimiglia, E. (2026). In vitro comparison of stone retropulsion between the new holmium:yttrium aluminium garnet laser generator with a very low peak power pulse modulation mode (Magneto technology) and the thulium fibre laser. BJU International, 137(2), 306-311. https://doi.org/10.1111/bju.70083
Vancouver
Folcia A, Perri D, Villa L, Albanese N, Robesti D, Corsini C, et al. In vitro comparison of stone retropulsion between the new holmium:yttrium aluminium garnet laser generator with a very low peak power pulse modulation mode (Magneto technology) and the thulium fibre laser. BJU International. 2026;137(2):306-311. doi:10.1111/bju.70083
BibTeX
@article{Folcia2026,
author = {Folcia, Andrea and Perri, Davide and Villa, Luca and Albanese, Nicole and Robesti, Daniele and Corsini, Christian and Bozzini, Giorgio and Kartalas Goumas, Ioannis and Hota, Thiago and Montorsi, Francesco and Salonia, Andrea and Traxer, Olivier and Ventimiglia, Eugenio},
title = {In vitro comparison of stone retropulsion between the new holmium:yttrium aluminium garnet laser generator with a very low peak power pulse modulation mode (Magneto technology) and the thulium fibre laser},
journal = {BJU International},
year = {2026},
volume = {137},
number = {2},
pages = {306--311},
doi = {10.1111/bju.70083}
}
EUVIMED