In Pursuit of the Optimal Dusting Settings with the Thulium Fiber Laser: An In Vitro Assessment
Journal of Endourology · 2023
DOI: 10.1089/end.2023.0168 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Endourology, 2023. Verfasst von Francois Soto-Palou und 6 weiteren Autor:innen. Beteiligte Einrichtung: Division of Urology, Department of Surgery, Duke University Medical Center, Durham, North Carolina, USA. Hauptthemen laut MEDLINE: Lithotripsy, Laser, Urinary Calculi, Kidney Calculi, Lasers, Solid-State. Die Arbeit wurde bisher 17-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIDDK NIH HHS.
Abstract (Originalsprache)
Objective
Low energy and high frequency settings are used in stone dusting for holmium lasers. Such settings may not be optimal for thulium fiber laser (TFL). With the seemingly endless combination of settings, we aim to provide guidance to the practicing urologists and assess the efficiency of the TFL platform in an automated in vitro “dusting model.” Materials/Methods: Three experimental setups were designed to investigate stone dusting produced by an IPG Photonics TLR-50 W TFL system using 200 μm fiber and soft BegoStone phantoms. The most popular 10 and 20 W dusting settings among endourologist familiar with TFL were evaluated. We directly compared short pulse (SP) vs long pulse (LP) mode using various combinations of pulse energy (E p ) and pulse frequency (F). Thereafter, we tested the 10 and 20 W settings and compared them among each other to elucidate the most efficient settings at each power. Treatments were performed under the same total laser energy delivered to the stone at four different standoff distances (SDs) with a clinically relevant scanning speed of either 1 or 2 mm/sec. Ablation volumes were quantified by optical coherence tomography to assess stone dusting efficiency. Fragment size after ablation at different pulse energies was evaluated by sieving and evaluating under a microscope after treatment.
Results
Overall, SP provided greater ablation volume when compared with LP. Our dusting efficiency model demonstrated that the maximum stone ablation was achieved at the combination of high energy/low frequency settings ( p < 0.005) and at a SD of 0.2 mm. At all tested pulse energies, no stone phantoms were broken into fragments >1 mm.
Conclusions
During stone dusting with TFL, SP offers superior ablation to LP settings. Optimal dusting at clinically relevant scanning speeds of 1 and 2 mm/sec occurs at high energy/low frequency settings. Thulium lithotripsy with high E p does not result in increased fragment size.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Division of Urology, Department of Surgery, Duke University Medical Center, Durham, North Carolina, USA.
- Department of Mechanical Engineering and Materials Science, Duke University, Durham, North Carolina, USA.
ORCID-Profile: Francois Soto-Palou
Förderung
- NIDDK NIH HHS
Zitieren
APA
Soto-Palou, F., Chen, J., Medairos, R., Zhong, P., Antonelli, J., Preminger, G. M., & Lipkin, M. E. (2023). In Pursuit of the Optimal Dusting Settings with the Thulium Fiber Laser: An In Vitro Assessment. Journal of Endourology. https://doi.org/10.1089/end.2023.0168
Vancouver
Soto-Palou F, Chen J, Medairos R, Zhong P, Antonelli J, Preminger GM, et al. In Pursuit of the Optimal Dusting Settings with the Thulium Fiber Laser: An In Vitro Assessment. Journal of Endourology. 2023. doi:10.1089/end.2023.0168
BibTeX
@article{SotoPalou2023,
author = {Soto-Palou, Francois and Chen, Junqin and Medairos, Robert and Zhong, Pei and Antonelli, Jodi and Preminger, Glenn M and Lipkin, Michael E},
title = {In Pursuit of the Optimal Dusting Settings with the Thulium Fiber Laser: An In Vitro Assessment},
journal = {Journal of Endourology},
year = {2023},
doi = {10.1089/end.2023.0168}
}
EUVIMED