Normative prospective data on automatically quantified retinal morphology correlated to retinal function in healthy ageing eyes by two microperimetry devices
Acta Ophthalmologica · 2025-06 · 103 (4) · 423-431
DOI: 10.1111/aos.17434 ↗Kurz gefasst
Fachartikel in der Zeitschrift Acta Ophthalmologica, online erschienen am 27.12.2024. Verfasst von Klaudia Birner und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology and Optometry, Medical University of Vienna, Vienna, Austria. Hauptthemen laut MEDLINE: Visual Field Tests, Tomography, Optical Coherence, Visual Fields, Retinal Ganglion Cells. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 46 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
The relationship between retinal morphology, as assessed by optical coherence tomography (OCT), and retinal function in microperimetry (MP) has not been well studied, despite its increasing importance as an essential functional endpoint for clinical trials and emerging therapies in retinal diseases. Normative databases of healthy ageing eyes are largely missing from literature.
Methods
Healthy subjects above 50 years were examined using two MP devices, MP‐3 (NIDEK) and MAIA (iCare). An identical grid, encompassing 45 stimuli was used for retinal sensitivity (RS) assessment. Deep‐learning‐based algorithms performed automated segmentation of ellipsoid zone (EZ), outer nuclear layer (ONL), ganglion cell layer (GCL) and retinal nerve fibre layer (RNFL) from OCT volumes (Spectralis, Heidelberg). Pointwise co‐registration between each MP stimulus and corresponding location on OCT was performed via registration algorithm. Effect of age, eccentricity and layer thickness on RS was assessed by mixed effect models.
Results
Three thousand six hundred stimuli from twenty eyes of twenty patients were included. Mean patient age was 68.9 ± 10.9 years. Mean RS for the first exam was 28.65 ± 2.49 dB and 25.5 ± 2.81 dB for MP‐3 and MAIA, respectively. Increased EZ thickness, ONL thickness and GCL thickness were significantly correlated with increased RS (all p < 0.001). Univariate models showed lower RS with advanced age and higher eccentricity (both p < 0.05).
Conclusion
This work provides reference values for healthy age‐related EZ and ONL‐thicknesses without impairment of visual function and evidence for RS decrease with eccentricity and increasing age. This information is crucial for interpretation of future clinical trials in disease.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Ophthalmology and Optometry, Medical University of Vienna, Vienna, Austria.
- Center for Medical Data Science, Institute of Medical Statistics, Medical University of Vienna, Vienna, Austria.
- Aritificial Intelligence Institute, Center for Medical Data Science, Medical University of Vienna, Vienna, Austria.
ORCID-Profile: Klaudia Birner, Johannes Schrittwieser, Ursula Schmidt-Erfurth, Gregor S Reiter
Zitieren
APA
Birner, K., Coulibaly, L. M., Schrittwieser, J., Steiner, I., Mohamed, H., Schürer-Waldheim, S., Gumpinger, M., Bogunovic, H., Schmidt-Erfurth, U., & Reiter, G. S. (2025). Normative prospective data on automatically quantified retinal morphology correlated to retinal function in healthy ageing eyes by two microperimetry devices. Acta Ophthalmologica, 103(4), 423-431. https://doi.org/10.1111/aos.17434
Vancouver
Birner K, Coulibaly LM, Schrittwieser J, Steiner I, Mohamed H, Schürer-Waldheim S, et al. Normative prospective data on automatically quantified retinal morphology correlated to retinal function in healthy ageing eyes by two microperimetry devices. Acta Ophthalmologica. 2025;103(4):423-431. doi:10.1111/aos.17434
BibTeX
@article{Birner2025,
author = {Birner, Klaudia and Coulibaly, Leonard M and Schrittwieser, Johannes and Steiner, Irene and Mohamed, Hamza and Schürer-Waldheim, Simon and Gumpinger, Markus and Bogunovic, Hrvoje and Schmidt-Erfurth, Ursula and Reiter, Gregor S},
title = {Normative prospective data on automatically quantified retinal morphology correlated to retinal function in healthy ageing eyes by two microperimetry devices},
journal = {Acta Ophthalmologica},
year = {2025},
volume = {103},
number = {4},
pages = {423--431},
doi = {10.1111/aos.17434}
}
EUVIMED