A 3D Print‐Based Experimental Framework for the Assessment of Haemocompatibility in Rotodynamic Blood Pumps

Aldo J Suria iD, Niccolò Crescenti, Benjamin Torner iD, Daniel Zimpfer, Marcus Granegger iD

Artificial Organs · 2026-11-13

DOI: 10.1111/aor.70285 ↗

At a glance

Journal article in Artificial Organs, published online on 2026-11-13. Written by Aldo J Suria and 4 co-authors. Institution: Department of Cardiac and Thoracic Aortic Surgery, Medical University of Vienna, Vienna, Austria. 36 references. The full text is freely available (open access, CC BY 4.0). Funded by: Austrian Science Fund.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Objective

Variability of 3D‐printed rotodynamic blood pumps (RBP) indicators is a crucial factor in haemocompatibility evaluation. Thus, the aim is to investigate the reproducibility of a 3D printed‐based framework for further use on systematic haemocompatible studies when changing key impeller parameters in RBPs.

Methods

Reproducibility was assessed in terms of standard deviation (SD), of the normalized index of haemolysis (NIH) and generated pressure head between 3 RBPs. A conventionally manufactured benchmark RBP, one 3D printed (3DP) exact replica, (), and one 3DP with a modification () ( n = 7). In vitro haemolysis experiments were carried out. Newly 3DP impellers were used for each experiment to study the manufacturing reproducibility. Moreover, pressure head‐flow rate curve analysis and computational fluid dynamic (CFD) simulations were performed to support the findings.

Results

The mean NIH ± SD in mg/100 L for the 3 RBPs was 4.25 ± 1.95 for the benchmark, 6.40 ± 2.68 for the exact replica, and 7.75 ± 1.87 for the modified version. Only a significant difference ( p < 0.05) between the benchmark and modified version was found. Conversely, no strong statistical differences were found between the 3DP RBPs. Furthermore, SD values of pressure heads were similar between all the groups. The resulting damage index for the 3DP geometries was 2.37 (·10 −4 ) and 2.32 (·10 −4 ) for the replica and modified version respectively. Suggesting a probable small effect when variating in free stream pumps.

Conclusions

Same NIH and pressure head SD range has shown the viability of the 3DP experimental framework, proposing an economic and rapid path for systematic haemocompatibility studies of parameters in RBPs.

Subjects

Author keywords

3D printinghaemocompatibilityhaemolysisrotodynamic blood pump

Authors and institutions

  • Department of Cardiac and Thoracic Aortic Surgery, Medical University of Vienna, Vienna, Austria.

ORCID profiles: Aldo J Suria, Benjamin Torner, Marcus Granegger

Funding

  • Austrian Science Fund

Cite

APA

Suria, A. J., Crescenti, N., Torner, B., Zimpfer, D., & Granegger, M. (2026). A 3D Print‐Based Experimental Framework for the Assessment of Haemocompatibility in Rotodynamic Blood Pumps. Artificial Organs. https://doi.org/10.1111/aor.70285

Vancouver

Suria AJ, Crescenti N, Torner B, Zimpfer D, Granegger M. A 3D Print‐Based Experimental Framework for the Assessment of Haemocompatibility in Rotodynamic Blood Pumps. Artificial Organs. 2026. doi:10.1111/aor.70285

BibTeX
@article{Suria2026,
  author = {Suria, Aldo J and Crescenti, Niccolò and Torner, Benjamin and Zimpfer, Daniel and Granegger, Marcus},
  title = {A 3D Print‐Based Experimental Framework for the Assessment of Haemocompatibility in Rotodynamic Blood Pumps},
  journal = {Artificial Organs},
  year = {2026},
  doi = {10.1111/aor.70285}
}
Start for free →