A 3D Print‐Based Experimental Framework for the Assessment of Haemocompatibility in Rotodynamic Blood Pumps
Artificial Organs · 2026-11-13
DOI: 10.1111/aor.70285 ↗At a glance
Journal article in Artificial Organs, published online on 2026-11-13. Written by Aldo J Suria and 4 co-authors. Institution: Department of Cardiac and Thoracic Aortic Surgery, Medical University of Vienna, Vienna, Austria. 36 references. The full text is freely available (open access, CC BY 4.0). Funded by: Austrian Science Fund.
Abstract (original language)
Objective
Variability of 3D‐printed rotodynamic blood pumps (RBP) indicators is a crucial factor in haemocompatibility evaluation. Thus, the aim is to investigate the reproducibility of a 3D printed‐based framework for further use on systematic haemocompatible studies when changing key impeller parameters in RBPs.
Methods
Reproducibility was assessed in terms of standard deviation (SD), of the normalized index of haemolysis (NIH) and generated pressure head between 3 RBPs. A conventionally manufactured benchmark RBP, one 3D printed (3DP) exact replica, (), and one 3DP with a modification () ( n = 7). In vitro haemolysis experiments were carried out. Newly 3DP impellers were used for each experiment to study the manufacturing reproducibility. Moreover, pressure head‐flow rate curve analysis and computational fluid dynamic (CFD) simulations were performed to support the findings.
Results
The mean NIH ± SD in mg/100 L for the 3 RBPs was 4.25 ± 1.95 for the benchmark, 6.40 ± 2.68 for the exact replica, and 7.75 ± 1.87 for the modified version. Only a significant difference ( p < 0.05) between the benchmark and modified version was found. Conversely, no strong statistical differences were found between the 3DP RBPs. Furthermore, SD values of pressure heads were similar between all the groups. The resulting damage index for the 3DP geometries was 2.37 (·10 −4 ) and 2.32 (·10 −4 ) for the replica and modified version respectively. Suggesting a probable small effect when variating in free stream pumps.
Conclusions
Same NIH and pressure head SD range has shown the viability of the 3DP experimental framework, proposing an economic and rapid path for systematic haemocompatibility studies of parameters in RBPs.
Subjects
Author keywords
Authors and institutions
- Department of Cardiac and Thoracic Aortic Surgery, Medical University of Vienna, Vienna, Austria.
ORCID profiles: Aldo J Suria, Benjamin Torner, Marcus Granegger
Funding
- Austrian Science Fund
Cite
APA
Suria, A. J., Crescenti, N., Torner, B., Zimpfer, D., & Granegger, M. (2026). A 3D Print‐Based Experimental Framework for the Assessment of Haemocompatibility in Rotodynamic Blood Pumps. Artificial Organs. https://doi.org/10.1111/aor.70285
Vancouver
Suria AJ, Crescenti N, Torner B, Zimpfer D, Granegger M. A 3D Print‐Based Experimental Framework for the Assessment of Haemocompatibility in Rotodynamic Blood Pumps. Artificial Organs. 2026. doi:10.1111/aor.70285
BibTeX
@article{Suria2026,
author = {Suria, Aldo J and Crescenti, Niccolò and Torner, Benjamin and Zimpfer, Daniel and Granegger, Marcus},
title = {A 3D Print‐Based Experimental Framework for the Assessment of Haemocompatibility in Rotodynamic Blood Pumps},
journal = {Artificial Organs},
year = {2026},
doi = {10.1111/aor.70285}
}
EUVIMED