A 3D Print‐Based Experimental Framework for the Assessment of Haemocompatibility in Rotodynamic Blood Pumps
Artificial Organs · 13.11.2026
DOI: 10.1111/aor.70285 ↗Kurz gefasst
Fachartikel in der Zeitschrift Artificial Organs, online erschienen am 13.11.2026. Verfasst von Aldo J Suria und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiac and Thoracic Aortic Surgery, Medical University of Vienna, Vienna, Austria. Die Arbeit nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Austrian Science Fund.
Abstract (Originalsprache)
Objective
Variability of 3D‐printed rotodynamic blood pumps (RBP) indicators is a crucial factor in haemocompatibility evaluation. Thus, the aim is to investigate the reproducibility of a 3D printed‐based framework for further use on systematic haemocompatible studies when changing key impeller parameters in RBPs.
Methods
Reproducibility was assessed in terms of standard deviation (SD), of the normalized index of haemolysis (NIH) and generated pressure head between 3 RBPs. A conventionally manufactured benchmark RBP, one 3D printed (3DP) exact replica, (), and one 3DP with a modification () ( n = 7). In vitro haemolysis experiments were carried out. Newly 3DP impellers were used for each experiment to study the manufacturing reproducibility. Moreover, pressure head‐flow rate curve analysis and computational fluid dynamic (CFD) simulations were performed to support the findings.
Results
The mean NIH ± SD in mg/100 L for the 3 RBPs was 4.25 ± 1.95 for the benchmark, 6.40 ± 2.68 for the exact replica, and 7.75 ± 1.87 for the modified version. Only a significant difference ( p < 0.05) between the benchmark and modified version was found. Conversely, no strong statistical differences were found between the 3DP RBPs. Furthermore, SD values of pressure heads were similar between all the groups. The resulting damage index for the 3DP geometries was 2.37 (·10 −4 ) and 2.32 (·10 −4 ) for the replica and modified version respectively. Suggesting a probable small effect when variating in free stream pumps.
Conclusions
Same NIH and pressure head SD range has shown the viability of the 3DP experimental framework, proposing an economic and rapid path for systematic haemocompatibility studies of parameters in RBPs.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiac and Thoracic Aortic Surgery, Medical University of Vienna, Vienna, Austria.
ORCID-Profile: Aldo J Suria, Benjamin Torner, Marcus Granegger
Förderung
- Austrian Science Fund
Zitieren
APA
Suria, A. J., Crescenti, N., Torner, B., Zimpfer, D., & Granegger, M. (2026). A 3D Print‐Based Experimental Framework for the Assessment of Haemocompatibility in Rotodynamic Blood Pumps. Artificial Organs. https://doi.org/10.1111/aor.70285
Vancouver
Suria AJ, Crescenti N, Torner B, Zimpfer D, Granegger M. A 3D Print‐Based Experimental Framework for the Assessment of Haemocompatibility in Rotodynamic Blood Pumps. Artificial Organs. 2026. doi:10.1111/aor.70285
BibTeX
@article{Suria2026,
author = {Suria, Aldo J and Crescenti, Niccolò and Torner, Benjamin and Zimpfer, Daniel and Granegger, Marcus},
title = {A 3D Print‐Based Experimental Framework for the Assessment of Haemocompatibility in Rotodynamic Blood Pumps},
journal = {Artificial Organs},
year = {2026},
doi = {10.1111/aor.70285}
}
EUVIMED