Geometry‐Related Technical Complications After LVAD Implantation: Impact on Device Dysfunction and Clinical Outcomes
Artificial Organs · 2026-11 · 50 (11) · 1605-1615
DOI: 10.1111/aor.70204 ↗Kurz gefasst
Fachartikel in der Zeitschrift Artificial Organs, online erschienen am 13.08.2026. Verfasst von Ali Saad Merzah und 17 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiac-, Thoracic-, Transplantation and Vascular Surgery, Hannover Medical School, Hannover, Germany. Hauptthemen laut MEDLINE: Heart-Assist Devices, Prosthesis Failure, Heart Failure, Prosthesis Implantation. Die Arbeit nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Geometry of LVAD implantation is a determinant of pump performance and hemocompatibility. Despite technically successful operations, avoidable geometric errors may produce characteristic imaging patterns and clinically relevant device dysfunction.
Objectives
To determine the frequency, diagnostic pathway, and clinical impact of preventable geometry‐related failure modes after HeartMate 3 implantation in a single‐center referral cohort, specifically inflow cannula malposition, outflow graft kinking/excessive length, and aortic anastomotic stenosis.
Methods
Retrospective, single‐center analysis of 21 consecutive HeartMate 3 recipients implanted between 2016 and 2025 at external centers and subsequently evaluated in our department. All patients underwent CT/CTA when device dysfunction was suspected. Failure modes were predefined and adjudicated using standardized imaging criteria. CT/CTA was the primary modality used to confirm geometric failure modes following triggers such as low‐flow alarms, hemolysis, and abnormal chest radiography.
Results
Among 21 HeartMate 3 recipients, inflow cannula malposition was identified in 7 patients, outflow graft kinking/excessive length in 12, and aortic anastomotic stenosis in 2.
Conclusions
In this single‐center referral cohort, geometric failure modes were identified in 15 of 21 patients (71.4%) after HeartMate 3 implantation and showed reproducible CT/CTA signatures. A standardized imaging‐driven diagnostic pathway may improve recognition of correctable geometric problems in patients referred for suspected device dysfunction; given the referral‐based design, these frequencies should not be interpreted as the true incidence in an unselected HeartMate 3 population.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiac-, Thoracic-, Transplantation and Vascular Surgery, Hannover Medical School, Hannover, Germany.
- Duisburg Cardiac Surgery Center, Duisburg, Germany.
- Kyiv Heart Institute, Kyiv, Ukraine.
- Department of Anesthesiology and Intensive Care Medicine, University Hospital of Cologne, Cologne, Germany.
- Department of Cardiac Surgery, Rostock Heart Center, University Medical Center Rostock, Rostock, Germany.
ORCID-Profile: Ali Saad Merzah, Ezin Deniz, Sara Knigge, Jasmin Hanke, Mohamad Dia, Chaikovska Sofia, Todurov Borys, Saeed Torabi, Günes Dogan
Zitieren
APA
Merzah, A. S., Deniz, E., Knigge, S., Hanke, J., Zubarevich, A., Dia, M., Sofia, C., Borys, T., Torabi, S., Schmack, B., Weymann, A., Darbaidze, Z., Haunschild, J., Etz, C. D., Biewer, M., Ruhparwar, A., Schmitto, J. D., & Dogan, G. (2026). Geometry‐Related Technical Complications After LVAD Implantation: Impact on Device Dysfunction and Clinical Outcomes. Artificial Organs, 50(11), 1605-1615. https://doi.org/10.1111/aor.70204
Vancouver
Merzah AS, Deniz E, Knigge S, Hanke J, Zubarevich A, Dia M, et al. Geometry‐Related Technical Complications After LVAD Implantation: Impact on Device Dysfunction and Clinical Outcomes. Artificial Organs. 2026;50(11):1605-1615. doi:10.1111/aor.70204
BibTeX
@article{Merzah2026,
author = {Merzah, Ali Saad and Deniz, Ezin and Knigge, Sara and Hanke, Jasmin and Zubarevich, Alina and Dia, Mohamad and Sofia, Chaikovska and Borys, Todurov and Torabi, Saeed and Schmack, Bastian and Weymann, Alexander and Darbaidze, Zurab and Haunschild, Josephina and Etz, Christian D and Biewer, Max and Ruhparwar, Arjang and Schmitto, Jan D and Dogan, Günes},
title = {Geometry‐Related Technical Complications After LVAD Implantation: Impact on Device Dysfunction and Clinical Outcomes},
journal = {Artificial Organs},
year = {2026},
volume = {50},
number = {11},
pages = {1605--1615},
doi = {10.1111/aor.70204}
}
EUVIMED