EUVIMEDEUVIMED entdecken
Crossref

Adverse Events and Failure Modes Related to Impella RP/RP Flex: Insights From FDA MAUDE Database

Kalyan R Chitturi, Ryan Wallace iD, Marta Lorente-Ros, Ilan Merdler iD, Abhishek Chaturvedi, Brian C Case, Hayder D Hashim, Itsik Ben-Dor, Toby Rogers, Ron Waksman iD

Catheterization and Cardiovascular Interventions · 2025

DOI: 10.1002/ccd.31333 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Catheterization and Cardiovascular Interventions, 2025. Verfasst von Kalyan R Chitturi und 9 weiteren Autor:innen. Beteiligte Einrichtung: Section of Interventional Cardiology, MedStar Washington Hospital Center, Washington, DC, USA. Hauptthemen laut MEDLINE: Heart-Assist Devices, United States Food and Drug Administration, Prosthesis Failure, Ventricular Function, Right. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 4 Literaturangaben. Gefördert von: The authors received no specific funding for this work..

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Impella RP and RP Flex can provide right ventricular (RV) support in numerous clinical situations. Notably, the redesigned RP Flex allows for an internal jugular approach compared to the femoral venous approach with RP.

Aims

We investigated the type and frequency of device‐related complications as reported in the US Food and Drug Administration (FDA) Manufacturer and User Facility Experience (MAUDE) for both Impella RP and RP Flex to see if there are any differences.

Methods

The MAUDE database was accessed from January 1, 2019 to April 30, 2024, yielding 206 reports pertaining to Impella RP and 64 to RP Flex. Reports were reviewed for device indications, mode of device failure, and device‐related complications.

Results

The most common indication for Impella RP placement was acute myocardial infarction with RV failure. The most commonly reported failure mode was pump failure. Death was reported in 8.5% and vascular complications in 3.7% of the reports. The most reported device‐related complication was hemolysis (18.5%), followed by bleeding requiring transfusion (14.1%), with new dialysis occurring in 3.3% of patients.

Conclusions

Our analysis of the MAUDE database showed that device delivery issues, bleeding, and vascular complications accounted for a smaller proportion of reports with Impella RP Flex than with its predecessor. However, device‐related thrombosis and pump malfunction accounted for a higher proportion of reports with RP Flex. These data do not reflect the overall incidence of complications, but instead highlight the need for physicians and other healthcare providers to return devices to the manufacturer for inspection when a failure occurs.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Heart-Assist DevicesUnited States Food and Drug AdministrationProsthesis FailureVentricular Function, RightProsthesis ImplantationHeart FailureHumansUnited StatesDatabases, FactualProsthesis DesignRisk FactorsTreatment OutcomeTime FactorsRisk AssessmentHemolysisHemorrhageMaleFemaleRetrospective StudiesMiddle Aged

Stichwörter der Autor:innen

adverse eventscardiogenic shockmicroaxial flow pumpright heart failureright ventricular assist device

Autor:innen und Institutionen

  • Section of Interventional Cardiology, MedStar Washington Hospital Center, Washington, DC, USA.
  • Department of Cardiology, MedStar Washington Hospital Center, Washington, DC, USA.

ORCID-Profile: Ryan Wallace, Ilan Merdler, Ron Waksman

Förderung

  • The authors received no specific funding for this work.

Zitieren

APA

Chitturi, K. R., Wallace, R., Lorente-Ros, M., Merdler, I., Chaturvedi, A., Case, B. C., Hashim, H. D., Ben-Dor, I., Rogers, T., & Waksman, R. (2025). Adverse Events and Failure Modes Related to Impella RP/RP Flex: Insights From FDA MAUDE Database. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.31333

Vancouver

Chitturi KR, Wallace R, Lorente-Ros M, Merdler I, Chaturvedi A, Case BC, et al. Adverse Events and Failure Modes Related to Impella RP/RP Flex: Insights From FDA MAUDE Database. Catheterization and Cardiovascular Interventions. 2025. doi:10.1002/ccd.31333

BibTeX
@article{Chitturi2025,
  author = {Chitturi, Kalyan R and Wallace, Ryan and Lorente-Ros, Marta and Merdler, Ilan and Chaturvedi, Abhishek and Case, Brian C and Hashim, Hayder D and Ben-Dor, Itsik and Rogers, Toby and Waksman, Ron},
  title = {Adverse Events and Failure Modes Related to Impella RP/RP Flex: Insights From FDA MAUDE Database},
  journal = {Catheterization and Cardiovascular Interventions},
  year = {2025},
  doi = {10.1002/ccd.31333}
}
Kostenlos starten →