Experimental evaluation of an artificial anal sphincter based on biomechanical compatibility

Minghui Wang iD, Wei Zhou, Yunlong Liu, Hongliu Yu

Artificial Organs · 2025

DOI: 10.1111/aor.14863 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Artificial Organs, 2025. Verfasst von Minghui Wang und 3 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Rehabilitation Engineering and Technology, School of Health Science and Engineering, University of Shanghai for Science and Technology, …. Hauptthemen laut MEDLINE: Anal Canal, Artificial Organs, Fecal Incontinence, Prosthesis Design. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 27 Literaturangaben. Gefördert von: National Natural Science Foundation of China.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The artificial anal sphincter is a device used to treat patients with fecal incontinence who are unable to control their bowel movements on their own. Long‐term morphological changes in the tissue surrounding the artificial anal sphincter can cause biomechanical compatibility problems, which seriously affect the clinical application of the artificial anal sphincter.

Methods

In this paper, the superelasticity of shape memory alloys was utilized to design and fabricate a biomechanically compatible constant force clamping artificial anal sphincter. An in vitro simulation system was constructed to verify the effectiveness, safety, and constant force characteristics of the artificial anal sphincter.

Results

The experimental results demonstrated that the artificial anal sphincter could be effectively closed with no leakage of the liquid‐like intestinal contents, which are most likely to leak. The pressure of the artificial anal sphincter on the intestinal tube gradually increased and eventually became constant during closure, and the pressure value was always less than the intestinal blood supply pressure threshold.

Conclusions

In this paper, we designed an artificial anal sphincter based on biomechanical compatibility and the corresponding in vitro simulation experimental program and preliminarily verified the effectiveness, safety, and constant force characteristics of the artificial anal sphincter.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Anal CanalArtificial OrgansFecal IncontinenceProsthesis DesignHumansBiomechanical PhenomenaAlloysMaterials TestingPressure

Stichwörter der Autor:innen

artificial anal sphincterbiomechanical compatibilityconstant force clampingfecal incontinencein vitro experimentsignal acquisition

Substanzen

Alloys

Autor:innen und Institutionen

  • Institute of Rehabilitation Engineering and Technology, School of Health Science and Engineering, University of Shanghai for Science and Technology, Shanghai, China.
  • Shanghai Engineering Research Center of Assistive Devices, Shanghai, China.

ORCID-Profile: Minghui Wang

Förderung

  • National Natural Science Foundation of China

Zitieren

APA

Wang, M., Zhou, W., Liu, Y., & Yu, H. (2025). Experimental evaluation of an artificial anal sphincter based on biomechanical compatibility. Artificial Organs. https://doi.org/10.1111/aor.14863

Vancouver

Wang M, Zhou W, Liu Y, Yu H. Experimental evaluation of an artificial anal sphincter based on biomechanical compatibility. Artificial Organs. 2025. doi:10.1111/aor.14863

BibTeX
@article{Wang2025,
  author = {Wang, Minghui and Zhou, Wei and Liu, Yunlong and Yu, Hongliu},
  title = {Experimental evaluation of an artificial anal sphincter based on biomechanical compatibility},
  journal = {Artificial Organs},
  year = {2025},
  doi = {10.1111/aor.14863}
}
Kostenlos starten →