Design and optimization of a new integrated hip and knee prosthesis structure
Artificial Organs · 2024
DOI: 10.1111/aor.14667 ↗Kurz gefasst
Fachartikel in der Zeitschrift Artificial Organs, 2024. Verfasst von Shengli Luo und 3 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Rehabilitation Engineering and Technology, University of Shanghai for Science and Technology, Shanghai, China. Hauptthemen laut MEDLINE: Knee Prosthesis, Arthroplasty, Replacement, Knee. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 32 Literaturangaben. Gefördert von: National Natural Science Foundation of China.
Abstract (Originalsprache)
Background
Conventional hip disarticulation prostheses (HDPs) are passive devices with separate joint structures, limiting amputees' ability to control and resulting in abnormal gait patterns. This study introduces a new HDP integrating the hip and knee joints for amputees' natural gait.
Methods
The new HDP restores the physiological rotation center of the hip with a remote center of motion (RCM) structure, and simulates the knee motion with a four‐bar structure. Nonlinear programming was employed to optimize the hip‐knee joint structure. A hybrid multi‐objective drive structure with a series–parallel connection was also designed to ensure motion synergy between the hip and knee joints. Finally, a prototype of the prosthesis was tested using the HDP test system.
Results
The optimization results demonstrate that the new HDP accurately restores the rotation center of the femur in amputees, with the knee's instantaneous center of rotation (ICR) trajectory closely resembling that of the human knee (Pearson correlation coefficient is 0.999). The study shows that the new HDP achieves a motion reproduction accuracy of over 95% for the human hip joint at walking speeds above 1.5 km/h, 38% higher than conventional prosthesis. Similarly, at the same walking speed, the new HDP replicates the motion of the human knee at 82.89%, surpassing conventional prosthesis by 57.85%.
Conclusions
The new HDP restores symmetry and replicates synergistic movement in amputees' lower limbs, exhibiting superior movement characteristics compared to conventional prostheses. This innovative HDP has the potential to enhance the quality of life for amputees.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Institute of Rehabilitation Engineering and Technology, University of Shanghai for Science and Technology, Shanghai, China.
ORCID-Profile: Shengli Luo, Hongliu Yu
Förderung
- National Natural Science Foundation of China
Zitieren
APA
Luo, S., Shu, X., Zhu, H., & Yu, H. (2024). Design and optimization of a new integrated hip and knee prosthesis structure. Artificial Organs. https://doi.org/10.1111/aor.14667
Vancouver
Luo S, Shu X, Zhu H, Yu H. Design and optimization of a new integrated hip and knee prosthesis structure. Artificial Organs. 2024. doi:10.1111/aor.14667
BibTeX
@article{Luo2024,
author = {Luo, Shengli and Shu, Xiaolong and Zhu, Hexiang and Yu, Hongliu},
title = {Design and optimization of a new integrated hip and knee prosthesis structure},
journal = {Artificial Organs},
year = {2024},
doi = {10.1111/aor.14667}
}