In‐silico hemodynamic ramp testing of ventricular assist device implanted patients using acausal cardiovascular‐VAD modeling

Khunsha Mehmood iD, Munam Arshad, Ismail Lazoglu iD, Deniz Süha Küçükaksu, Vedat Bakuy

Artificial Organs · 2023

DOI: 10.1111/aor.14597 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Artificial Organs, 2023. Verfasst von Khunsha Mehmood und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechanical Engineering, Koç University, Istanbul, Turkey. Hauptthemen laut MEDLINE: Heart-Assist Devices, Cardiovascular System, Heart Failure. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 42 Literaturangaben. Gefördert von: Türkiye Bilimsel ve Teknolojik Araştırma Kurumu.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

While cardiovascular system and mechanical circulatory support devices are efficiently model the effect of disease and assistance, they can also lend valuable insights into clinical procedures. This study demonstrates the use of a CVS‐VAD model for an invasive procedure; hemodynamic ramp testing, in‐silico.

Methods

The CVS model is developed using validated models in literature, using Simscape™. An analytically derived pump model is calibrated for the HeartWare VAD. Dilated cardiomyopathy is used as an illustrative example of heart failure, and heart failure patients are created virtually by calibrating the model with requisite disease parameters obtained from published patient data. A clinically applied ramp study protocol is adopted whereby speed optimization is performed following clinically accepted hemodynamic normalization criteria. Trends in hemodynamic variables in response to pump speed increments are obtained. Optimal speed ranges are obtained for the three virtual patients based on target values of central venous pressure (CVP), pulmonary capillary wedge pressure (PCWP), cardiac output (CO), and mean arterial pressure (MAP) for hemodynamic stabilization.

Results

Appreciable speed changes in the mild case (300 rpm), slight changes in the moderate case (100 rpm), and no changes in the simulated severe case are possible.

Conclusion

The study demonstrates a novel application of cardiovascular modeling using an open‐source acausal model, which can be potentially beneficial for medical education and research.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Heart-Assist DevicesCardiovascular SystemHeart FailureHumansHemodynamicsCardiac Output

Stichwörter der Autor:innen

HeartWareSimscapeobject-oriented modelingoptimal hemodynamics

Autor:innen und Institutionen

  • Department of Mechanical Engineering, Koç University, Istanbul, Turkey.
  • Department of Cardiovascular Surgery, Başkent University Istanbul Health Application and Research Center, Istanbul, Turkey.

ORCID-Profile: Khunsha Mehmood, Ismail Lazoglu

Förderung

  • Türkiye Bilimsel ve Teknolojik Araştırma Kurumu

Zitieren

APA

Mehmood, K., Arshad, M., Lazoglu, I., Küçükaksu, D. S., & Bakuy, V. (2023). In‐silico hemodynamic ramp testing of ventricular assist device implanted patients using acausal cardiovascular‐VAD modeling. Artificial Organs. https://doi.org/10.1111/aor.14597

Vancouver

Mehmood K, Arshad M, Lazoglu I, Küçükaksu DS, Bakuy V. In‐silico hemodynamic ramp testing of ventricular assist device implanted patients using acausal cardiovascular‐VAD modeling. Artificial Organs. 2023. doi:10.1111/aor.14597

BibTeX
@article{Mehmood2023,
  author = {Mehmood, Khunsha and Arshad, Munam and Lazoglu, Ismail and Küçükaksu, Deniz Süha and Bakuy, Vedat},
  title = {In‐silico hemodynamic ramp testing of ventricular assist device implanted patients using acausal cardiovascular‐VAD modeling},
  journal = {Artificial Organs},
  year = {2023},
  doi = {10.1111/aor.14597}
}
Kostenlos starten →