Impact of an improved driveline management for HeartMate II and HeartMate 3 left ventricular assist devices
Artificial Organs · 2023-02 · 47 (2) · 387-395
DOI: 10.1111/aor.14426 ↗Kurz gefasst
Fachartikel in der Zeitschrift Artificial Organs, online erschienen am 02.11.2022. Verfasst von Masato Mutsuga und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiac Surgery, Nagoya University Graduate School of Medicine, Nagoya, Japan. Hauptthemen laut MEDLINE: Heart-Assist Devices, Heart Failure, Prosthesis-Related Infections. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 27 Literaturangaben.
Abstract (Originalsprache)
Background
We evaluated the impact of a standardized driveline care strategy, including a subfascial‐tunneling method and dressing protocol, on the incidence of driveline infection (DLI).
Methods
DLI data from all HeartMate II (HMII) and HeartMate 3 (HM3) patients (including exchange devices) were retrospectively collected between 2013 and 2021. The driveline subfascial‐tunneling method was altered in three steps (A: right direct; B: left triple, C: right triple), and the shower protocol was changed in two steps (A: with/without cover, B: with cover). Disinfection was individually tailored after changing the shower protocol. Complications associated with morbidity and mortality were evaluated for each modification.
Results
During the study period, 80 devices were implanted (HMII, n = 54; HM3, n = 26). The 8‐year incidence of DLI was 15% ( n = 8) in HMII patients and 0% in HM3 patients ( p = 0.039). DLI was not associated with hospital mortality. The modified dressing protocol and tunneling method was associated with a significantly better DLI incidence rate in comparison to the previous one: Protocol‐A ( n = 17), Protocol‐B ( n = 63), 35% vs 3% ( p = 0.0009), Method‐A ( n = 13), Method‐B ( n = 42), Method‐C ( n = 25), 46% vs 5% vs 0% ( p = 0.0001). The rete of freedom form DLI at 1, 2, and 3 years had also significant difference between groups: Protocol‐A and Protocol‐B, 80%, 54%, 54% vs 96%, 96%, 96%, respectively ( p < 0.0001), Method‐A, Method‐B and Method‐C, 76%, 44%, 44%, vs 94%, 94%, 94% vs 100%, 100%, respectively ( p < 0.0001).
Conclusions
A standardized triple driveline tunneling strategy and waterproof dressing protocol reduced driveline infection in HM3 patients to 0%.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiac Surgery, Nagoya University Graduate School of Medicine, Nagoya, Japan.
- Department of Cardiology, Nagoya University Graduate School of Medicine, Nagoya, Japan.
- Department of Biostatistics Section, Center for Advanced Medicine and Clinical Research, Nagoya University Graduate School of Medicine, Nagoya, Japan.
ORCID-Profile: Masato Mutsuga, Takahiro Okumura, Toru Kondo, Kazuki Nishida
Zitieren
APA
Mutsuga, M., Okumura, T., Morimoto, R., Kondo, T., Ito, H., Terazawa, S., Tokuda, Y., Narita, Y., Nishida, K., Murohara, T., & Usui, A. (2023). Impact of an improved driveline management for HeartMate II and HeartMate 3 left ventricular assist devices. Artificial Organs, 47(2), 387-395. https://doi.org/10.1111/aor.14426
Vancouver
Mutsuga M, Okumura T, Morimoto R, Kondo T, Ito H, Terazawa S, et al. Impact of an improved driveline management for HeartMate II and HeartMate 3 left ventricular assist devices. Artificial Organs. 2023;47(2):387-395. doi:10.1111/aor.14426
BibTeX
@article{Mutsuga2023,
author = {Mutsuga, Masato and Okumura, Takahiro and Morimoto, Ryota and Kondo, Toru and Ito, Hideki and Terazawa, Sachie and Tokuda, Yoshiyuki and Narita, Yuji and Nishida, Kazuki and Murohara, Toyoaki and Usui, Akihiko},
title = {Impact of an improved driveline management for HeartMate II and HeartMate 3 left ventricular assist devices},
journal = {Artificial Organs},
year = {2023},
volume = {47},
number = {2},
pages = {387--395},
doi = {10.1111/aor.14426}
}
EUVIMED