Kurz gefasst

Multizentrische Studie in der Zeitschrift European Journal of Cardio-Thoracic Surgery, online erschienen am 25.10.2025. Verfasst von Xianqiang Wang und 15 weiteren Autor:innen. Beteiligte Einrichtung: Centre of Cardiac Surgery in Adults, Fuwai Hospital, National Center for Cardiovascular Diseases, Chinese Academy of Medical Sciences and Peking Unio…. Hauptthemen laut MEDLINE: Heart-Assist Devices, Prosthesis-Related Infections, Heart Failure. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 18 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Beijing Outstanding Young Scientist Program.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

Driveline infection remains a major complication of long-term ventricular assist device therapy. Previous studies have associated the risk of infection with driveline design and physical properties. The CH-VAD is a fully magnetically levitated pump with an ultrathin (diameter 3.3 mm), highly flexible driveline designed to reduce infection risk. This study was conducted to assess the incidence, risk factors, and approaches to prevention and management of driveline infection in CH-VAD patients (NCT06878456).

Methods

All patients who underwent CH-VAD implantation between June 2017 and October 2024 at 9 centres were retrospectively analysed. Data related to clinical characteristics, surgical procedures, and driveline infection prevention and management details were collected, and risk factors were analysed.

Results

A total of 181 consecutive patients were included, with a median age of 56 years and 86% male. During a median support duration of 1.6 years, driveline infections occurred in 20 out of 181 patients (11.0%). Twenty-six events were observed, corresponding to a rate of 0.07 events per patient-year. Cumulative incidence of the first driveline infection was 2.8% at 6 months, 5.6% at 12 months, and 8.5% at 24 months. Frequent dressing changes were found to be significantly associated with increased driveline infection risk (HR 3.773, P = .022).

Conclusions

Patients treated with CH-VADs experienced an encouraging low rate of driveline infection, considered among the lowest reported for durable ventricular assist devices. These findings have implications for both clinical management and driveline design.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Heart-Assist DevicesProsthesis-Related InfectionsHeart FailureHumansMaleMiddle AgedFemaleRetrospective StudiesRisk FactorsAgedAdultIncidenceProsthesis Design

Stichwörter der Autor:innen

CH-VADadvanced heart failuredriveline infectionleft ventricular assist device

Autor:innen und Institutionen

  • Centre of Cardiac Surgery in Adults, Fuwai Hospital, National Center for Cardiovascular Diseases, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, 100037, China.
  • Department of Cardiothoracic Surgery, Zhejiang Provincial People's Hospital, Hangzhou, 310014, China.
  • Department of Cardiac Surgery, Sichuan Provincial People's Hospital, Chengdu, 610072, China.
  • Department of Cardiac Surgery, Asia Heart Hospital, Wuhan, 430022, China.
  • Department of Cardiovascular Surgery, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, 710061, China.
  • Department of Cardiovascular Surgery, Ruijin Hospital, Shanghai Jiaotong University School of Medicine, Shanghai, 200025, China.

ORCID-Profile: Yang Yan, Qiang Zhao, Shengshou Hu

Förderung

  • Beijing Outstanding Young Scientist Program

Zitieren

APA

Wang, X., Zhou, X., Chen, H., Zhang, H., Zhang, Y., Cui, Y., Huang, K., Hua, Z., Yan, Y., Zhou, M., Zhao, Q., Sun, X., Wang, C., Ge, Z., Dong, A., & Hu, S. (2025). Low Driveline Infection Rates in Patients With a Novel Fully Magnetically Levitated Ventricular Assist Device. European Journal of Cardio-Thoracic Surgery, 67(12). https://doi.org/10.1093/ejcts/ezaf365

Vancouver

Wang X, Zhou X, Chen H, Zhang H, Zhang Y, Cui Y, et al. Low Driveline Infection Rates in Patients With a Novel Fully Magnetically Levitated Ventricular Assist Device. European Journal of Cardio-Thoracic Surgery. 2025;67(12). doi:10.1093/ejcts/ezaf365

BibTeX
@article{Wang2025,
  author = {Wang, Xianqiang and Zhou, Xingtong and Chen, Haibo and Zhang, Haotian and Zhang, Yanming and Cui, Yong and Huang, Keli and Hua, Zhengdong and Yan, Yang and Zhou, Mi and Zhao, Qiang and Sun, Xiaoning and Wang, Chunsheng and Ge, Zhenwei and Dong, Aiqiang and Hu, Shengshou},
  title = {Low Driveline Infection Rates in Patients With a Novel Fully Magnetically Levitated Ventricular Assist Device},
  journal = {European Journal of Cardio-Thoracic Surgery},
  year = {2025},
  volume = {67},
  number = {12},
  doi = {10.1093/ejcts/ezaf365}
}
Kostenlos starten →