Accuracy of Apex Locator Integrated Endomotors in Estimating Working Length in the Presence of Endodontic Irrigants: An Ex Vivo Study
Australian Endodontic Journal · 2025
DOI: 10.1111/aej.70002 ↗Kurz gefasst
Fachartikel in der Zeitschrift Australian Endodontic Journal, 2025. Verfasst von Kavalipurapu Venkata Teja und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Conservative Dentistry and Endodontics, Mamata Institute of Dental Sciences, Hyderabad, Telangana, India. Hauptthemen laut MEDLINE: Root Canal Irrigants, Tooth Apex, Root Canal Preparation, Dental Pulp Cavity. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: GEBR. Brasseler GmbH & Co. KG, Trophagener Weg 25, 32657 Lemgo, Germany..
Abstract (Originalsprache)
Accurate working length determination is crucial for root canal treatment. This ex vivo study evaluated the accuracy of a standalone apex locator (Root ZX Mini) and two integrated apex locator‐endomotors (EnDrive and X‐Smart Pro+) under different intracanal conditions. Seventy‐two extracted single‐rooted mandibular premolars were decoronated and standardised. The actual canal length (AML) was determined microscopically, and electronic working lengths (EWL) obtained. Measurements were taken after canal patency and drying with paper points or following irrigation with saline, sodium hypochlorite, or ethylenediaminetetraacetic acid. All devices and intracanal conditions exhibited negative deviations, with EWLs shorter than AML, ranging from −0.15 to −0.50 mm, thus remaining within clinically acceptable limits. Significant differences were observed among devices and canal contents ( p < 0.05), with EnDrive showing the least deviation, particularly in the presence of sodium hypochlorite. These findings support the reliability of integrated apex locator‐endomotors, though measurement accuracy may be affected by irrigant type.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Conservative Dentistry and Endodontics, Mamata Institute of Dental Sciences, Hyderabad, Telangana, India.
- Department of Prosthodontics, Saveetha Dental College and Hospitals, Saveetha Institute of Medical and Technical Sciences, Saveetha University, Chennai, Tamil Nadu, India.
- Department of Conservative Dentistry and Endodontics, Saveetha Dental College and Hospitals, Saveetha Institute of Medical and Technical Sciences, Saveetha University, Chennai, Tamil Nadu, India.
- Adelaide Dental School, the University of Adelaide, Adelaide, South Australia, Australia.
ORCID-Profile: Kavalipurapu Venkata Teja, Pradeep Solete, Giampiero Rossi-Fedele
Förderung
- GEBR. Brasseler GmbH & Co. KG, Trophagener Weg 25, 32657 Lemgo, Germany.
Zitieren
APA
Teja, K. V., Vasundhara, K. A., Solete, P., & Rossi-Fedele, G. (2025). Accuracy of Apex Locator Integrated Endomotors in Estimating Working Length in the Presence of Endodontic Irrigants: An Ex Vivo Study. Australian Endodontic Journal. https://doi.org/10.1111/aej.70002
Vancouver
Teja KV, Vasundhara KA, Solete P, Rossi-Fedele G. Accuracy of Apex Locator Integrated Endomotors in Estimating Working Length in the Presence of Endodontic Irrigants: An Ex Vivo Study. Australian Endodontic Journal. 2025. doi:10.1111/aej.70002
BibTeX
@article{Teja2025,
author = {Teja, Kavalipurapu Venkata and Vasundhara, Kaligotla Apoorva and Solete, Pradeep and Rossi-Fedele, Giampiero},
title = {Accuracy of Apex Locator Integrated Endomotors in Estimating Working Length in the Presence of Endodontic Irrigants: An Ex Vivo Study},
journal = {Australian Endodontic Journal},
year = {2025},
doi = {10.1111/aej.70002}
}