Efficiency comparison between new single‐needle dialysis and conventional double‐needle dialysis and exploration of the optimal prescription parameters for single‐needle hemodialysis
Therapeutic Apheresis and Dialysis · 2025
DOI: 10.1111/1744-9987.70006 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Therapeutic Apheresis and Dialysis, 2025. Verfasst von Bin Zhao und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Nephrology, Beijing Haidian Hospital (Haidian Section of Pecking University Third Hospital), Beijing, China. Hauptthemen laut MEDLINE: Renal Dialysis, Kidney Failure, Chronic. Die Arbeit nennt 13 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Beijing Haidian Hospital important project.
Abstract (Originalsprache)
Objective
To compare the dialysis efficiency parameters between new single‐needle dialysis and conventional double‐needle dialysis and to explore the optimal prescription parameters for single‐needle hemodialysis.
Methods
This study involved a prospective, cross‐over design. Eight patients were randomly divided into two cohorts, which received the appropriate treatment. Relevant clinical data and dialysis‐related parameters were collected for further analysis.
Results
Each of the patients enrolled in this study underwent nine sessions of single‐needle dialysis and three sessions of double‐needle dialysis. The average blood flow after 4, 4.5, and 5 h of single‐needle dialysis was significantly lower ( p < 0.001) than that with double‐needle dialysis for the equivalent times. The online Kt/V value measured after 4 h of single‐needle dialysis was significantly lower than that with 4 h of double‐needle dialysis ( p < 0.001). When replacing the dialyzer with a high‐flux dialyzer (FX800) and extending the dialysis time to 5 h, there was no difference in the online Kt/V values ( p = 0.341). There was also no difference in the ultrafiltration volume between single‐needle and double‐needle dialysis for the equivalent times. The calculation of the Kt/V and urea reduction rate (URR) showed that there was no difference between 5 h of single‐needle dialysis and conventional double‐needle dialysis ( p > 0.05).
Conclusions
The new modified single‐needle system is safe and effective to support the dialysis treatment of maintenance hemodialysis patients. When adjusting the dialysis prescription to a high‐flux dialyzer and extending the dialysis time to 5 h, it can achieve dialysis efficiency equivalent to that of conventional 4 h double‐needle dialysis.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Nephrology, Beijing Haidian Hospital (Haidian Section of Pecking University Third Hospital), Beijing, China.
ORCID-Profile: Shen Zhan, Yuzhu Wang
Förderung
- Beijing Haidian Hospital important project
Zitieren
APA
Zhao, B., Zhan, S., Fu, G., Guo, S., Liu, L., Liu, W., Xiao, G., Wang, Y., & Zhang, L. (2025). Efficiency comparison between new single‐needle dialysis and conventional double‐needle dialysis and exploration of the optimal prescription parameters for single‐needle hemodialysis. Therapeutic Apheresis and Dialysis. https://doi.org/10.1111/1744-9987.70006
Vancouver
Zhao B, Zhan S, Fu G, Guo S, Liu L, Liu W, et al. Efficiency comparison between new single‐needle dialysis and conventional double‐needle dialysis and exploration of the optimal prescription parameters for single‐needle hemodialysis. Therapeutic Apheresis and Dialysis. 2025. doi:10.1111/1744-9987.70006
BibTeX
@article{Zhao2025,
author = {Zhao, Bin and Zhan, Shen and Fu, Gang and Guo, Shanshan and Liu, Lifang and Liu, Wei and Xiao, Guanghui and Wang, Yuzhu and Zhang, Lihong},
title = {Efficiency comparison between new single‐needle dialysis and conventional double‐needle dialysis and exploration of the optimal prescription parameters for single‐needle hemodialysis},
journal = {Therapeutic Apheresis and Dialysis},
year = {2025},
doi = {10.1111/1744-9987.70006}
}