EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Hadamard-Based Recursive Aperture Decoded Ultrasound Imaging (READI) With Estimated Motion-Compensated Compounding (EMC 2 ) Using Top-Orthogonal-to-Bottom-Electrode (TOBE) Arrays

Tyler Keith Henry, Darren Dahunsi, Randy Palamar, Negar Majidi, Ying Wan, Mohammad Rahim Sobhani, Afshin Kashani Ilkhechi, Roger Zemp

IEEE Transactions on Ultrasonics · 2026

DOI: 10.1109/tuson.2026.3704971 ↗

Kurz gefasst

Fachartikel in der Zeitschrift IEEE Transactions on Ultrasonics, 2026. Verfasst von Tyler Keith Henry und 7 weiteren Autor:innen. Hauptthemen laut MEDLINE: Image Processing, Computer-Assisted, Signal Processing, Computer-Assisted. Die Arbeit nennt 53 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NEI NIH HHS, NHLBI NIH HHS, Advanced Research Projects Agency for Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Hadamard matrix-based aperture encoding is a method for producing synthetic aperture datasets with high signal-to-noise ratios (SNRs). Recently, the pulse inversion capabilities of bias-sensitive top-orthogonal-to-bottom-electrode (TOBE) arrays have driven the development of multiple Hadamard-based sequences. These sequences produce high-quality static images but are sensitive to motion. This work introduces recursive aperture decoded ultrasound imaging (READI) and estimated motion-compensated compounding (EMC2), which look to reduce this sensitivity. READI is a novel decoding and beamforming technique for Hadamard aperture-encoded sequences that produces multiple low-resolution images from subsets of the full sequence. These READI images are less affected by motion and sum to form the complete high-resolution image. EMC2 describes the process of comparing these low-resolution images to estimate the underlying motion, then warping them to align before compounding. This produces a high-resolution image that is resilient to motion. READI with EMC2 applied to the TOBE-based fast orthogonal row-column electronic scanning (FORCES) sequence. It is shown to fully restore images corrupted by probe motion and to recover tissue speckle and boundaries in images of a beating heart phantom. READI low-resolution images by themselves are demonstrated to be a marked improvement over a sparse Hadamard scheme with the same transmit count, and are able to recover blood speckle at a flow rate of 42 cm/s.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Image Processing, Computer-AssistedSignal Processing, Computer-AssistedSignal-To-Noise RatioPhantoms, ImagingAlgorithmsUltrasonographyHumansMotion

Förderung

  • NEI NIH HHS
  • NHLBI NIH HHS
  • Advanced Research Projects Agency for Health
  • National Institutes of Health
  • Alberta Innovates
  • MITACS
  • CliniSonix Inc., NSERC
  • Alberta Cancer Foundation and the Mary Johnston Family Melanoma

Zitieren

APA

Henry, T. K., Dahunsi, D., Palamar, R., Majidi, N., Wan, Y., Sobhani, M. R., Kashani Ilkhechi, A., & Zemp, R. (2026). Hadamard-Based Recursive Aperture Decoded Ultrasound Imaging (READI) With Estimated Motion-Compensated Compounding (EMC 2 ) Using Top-Orthogonal-to-Bottom-Electrode (TOBE) Arrays. IEEE Transactions on Ultrasonics. https://doi.org/10.1109/tuson.2026.3704971

Vancouver

Henry TK, Dahunsi D, Palamar R, Majidi N, Wan Y, Sobhani MR, et al. Hadamard-Based Recursive Aperture Decoded Ultrasound Imaging (READI) With Estimated Motion-Compensated Compounding (EMC 2 ) Using Top-Orthogonal-to-Bottom-Electrode (TOBE) Arrays. IEEE Transactions on Ultrasonics. 2026. doi:10.1109/tuson.2026.3704971

BibTeX
@article{Henry2026,
  author = {Henry, Tyler Keith and Dahunsi, Darren and Palamar, Randy and Majidi, Negar and Wan, Ying and Sobhani, Mohammad Rahim and Kashani Ilkhechi, Afshin and Zemp, Roger},
  title = {Hadamard-Based Recursive Aperture Decoded Ultrasound Imaging (READI) With Estimated Motion-Compensated Compounding (EMC 2 ) Using Top-Orthogonal-to-Bottom-Electrode (TOBE) Arrays},
  journal = {IEEE Transactions on Ultrasonics},
  year = {2026},
  doi = {10.1109/tuson.2026.3704971}
}
Kostenlos starten →