EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

Analytic Timing Calculations and Timing Limits With Prompt Photons, High-Aspect-Ratio Crystals, and Complex TOF-Kernels in TOF-PET

Nicolaus Kratochwil, Emilie Roncali, Gerard Ariño-Estrada

IEEE Transactions on Radiation and Plasma Medical Sciences · 2026-11 · 10 (8) · 1276-1288

DOI: 10.1109/trpms.2026.3665464 ↗

Kurz gefasst

Fachartikel in der Zeitschrift IEEE Transactions on Radiation and Plasma Medical Sciences, 2026-11. Verfasst von Nicolaus Kratochwil und 2 weiteren Autor:innen. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 74 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: National Institute of Health.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Modeling the timing performance of light-based radiation detectors accurately is essential for optimizing time-of- flight positron emission tomography (TOF-PET). We present an analytic framework that combines existing models to predict the timing behavior of high-aspect ratio crystals, including contributions from prompt photons such as Cherenkov radiation. This framework is built on a closed-form solution for optical light transport, convolved with the photodetector response and photon production characteristics. Using conditional and joint probability distributions, we compute the first-photon arrival time distribution for hybrid detectors with scintillation and Cherenkov light. The detection time distribution is then self-convolved to derive the time delay spectra and three timing metrics are used to characterize complex TOF kernels. Additionally, we perform Cramér-Rao Lower Bound calculations with and without depth-of- interaction bias to evaluate the theoretical timing limits. Our analytic predictions align well with Monte Carlo simulations for BGO detectors under varying crystal thicknesses and single photon time resolution considering a digital photodetector. We show that the TOF kernel shape is significantly affected by prompt photon statistics, crystal thickness, scintillation yield, and photodetector properties resulting in distinct metric-dependent timing performance. The proposed model enables rapid timing predictions for polished crystals, with the calculation time of a detector configuration in under a second, allowing for comprehensive parametric studies. This makes it a powerful tool for guiding detector development in fast-timing applications.

Förderung

  • National Institute of Health

Zitieren

APA

Kratochwil, N., Roncali, E., & Ariño-Estrada, G. (2026). Analytic Timing Calculations and Timing Limits With Prompt Photons, High-Aspect-Ratio Crystals, and Complex TOF-Kernels in TOF-PET. IEEE Transactions on Radiation and Plasma Medical Sciences, 10(8), 1276-1288. https://doi.org/10.1109/trpms.2026.3665464

Vancouver

Kratochwil N, Roncali E, Ariño-Estrada G. Analytic Timing Calculations and Timing Limits With Prompt Photons, High-Aspect-Ratio Crystals, and Complex TOF-Kernels in TOF-PET. IEEE Transactions on Radiation and Plasma Medical Sciences. 2026;10(8):1276-1288. doi:10.1109/trpms.2026.3665464

BibTeX
@article{Kratochwil2026,
  author = {Kratochwil, Nicolaus and Roncali, Emilie and Ariño-Estrada, Gerard},
  title = {Analytic Timing Calculations and Timing Limits With Prompt Photons, High-Aspect-Ratio Crystals, and Complex TOF-Kernels in TOF-PET},
  journal = {IEEE Transactions on Radiation and Plasma Medical Sciences},
  year = {2026},
  volume = {10},
  number = {8},
  pages = {1276--1288},
  doi = {10.1109/trpms.2026.3665464}
}