Performance Characteristics Of LYSO Scintillation Crystals Based On 15cm (3 Ring) TOF PET/CT
IOSR Journal of Nursing and health Science · 2026-02 · 15 (2) · 13-22
DOI: 10.9790/1959-1502031322 ↗Kurz gefasst
Fachartikel in der Zeitschrift IOSR Journal of Nursing and health Science, 2026-02. Verfasst von Rasool Kattubadi und 2 weiteren Autor:innen.
Abstract (Originalsprache)
Background
This paper represents the National Electrical Manufacturers Association (NEMA) system performance of the Discovery MI 3-ring PET/CT (GE Healthcare) installed in advanced cardiac unit, King Abdulaziz Hospital (KAH), Advanced cardiac unit, Riyadh. This time-of-flight (TOF) PET camera is based on lutetium-based scintillation crystals. LYSO, Lutetium–yttrium oxyorthosilicate is an inorganic chemical compound with main use as a scintillator crystal for gamma radiation detection.
Methods
The NEMA NU2-2018 standard was used to evaluate spatial resolution, sensitivity, image quality (IQ) and count rate curves of the system. Timing and energy resolution were determined.
Results
Full width at half maximum (FWHM) of spatial resolution in radial, tangential and axial direction was 3.84, 3.84 and 4.18 mm at 1 cm; 4.66, 4.08 and 4.91 mm at 10 cm; and 7.15, 5.31 and 5.47 mm at 20 cm from the centre of the field of view (FOV) for the VPHD reconstruction. For non-TOF ordered subset expectation maximization (OSEM) reconstruction without point spread function (PSF) correction, FWHM was 3.87, 3.69 and 4.15 mm at 1 cm; 4.80, 3.81 and 4.87 mm at 10 cm; and 7.38, 4.16 and 3.98 mm at 20 cm. Sensitivity was 7.12 cps/kBq at the centre of the FOV and 6.84 cps/kBq at 10-cm radial offset. Contrast recovery (CR) using the IQ phantom for the TOF VPFX reconstruction was 105.3, 119.7, 126.8 and 98.1% for the 10-, 13-, 17- and 22-mm radioactive spheres and 96.5 and 92.2% for the 28- and 37-mm non-radioactive spheres. Background variability (BV) was 17.5, 12.7, 9.0, 7.1, 5.5 and 4.5% for the 10-, 13-, 17-, 22-, 28- and 37-mm spheres. Lung error was 7.8%. Peak noise equivalent count rate (NECR) was 114.3 kcps at 22.20 kBq/ml with a scatter fraction of 42.10%. Maximum accuracy error was 3.88%. Coincidence timing resolution at 5.3 kBq/ml was 411.2 ps FWHM. Energy resolution was 9.94% FWHM. Observed that Q.Clear reconstruction significantly improved CR and reduced BV compared with OSEM while processing the data.
Conclusion
System sensitivity and NECR are lower and IQ phantom’s BV is higher compared with larger axial FOV (AFOV) scanners like the 4-ring discovery MI, as expected from the smaller solid angle of the 3-ring system. The other NEMA performance parameters are all comparable with those of the larger AFOV scanners.
Zitieren
APA
Kattubadi, R., Mohamedahmed, E., & Almajid, H. (2026). Performance Characteristics Of LYSO Scintillation Crystals Based On 15cm (3 Ring) TOF PET/CT. IOSR Journal of Nursing and health Science, 15(2), 13-22. https://doi.org/10.9790/1959-1502031322
Vancouver
Kattubadi R, Mohamedahmed E, Almajid H. Performance Characteristics Of LYSO Scintillation Crystals Based On 15cm (3 Ring) TOF PET/CT. IOSR Journal of Nursing and health Science. 2026;15(2):13-22. doi:10.9790/1959-1502031322
BibTeX
@article{Kattubadi2026,
author = {Kattubadi, Rasool and Mohamedahmed, Elmubarak and Almajid, Hamad},
title = {Performance Characteristics Of LYSO Scintillation Crystals Based On 15cm (3 Ring) TOF PET/CT},
journal = {IOSR Journal of Nursing and health Science},
year = {2026},
volume = {15},
number = {2},
pages = {13--22},
doi = {10.9790/1959-1502031322}
}
EUVIMED