Performance Evaluation of Single- and Double-Layer CZT Compton Cameras for 225 Ac Imaging

Gyohyeok Song iD, Biswajit Das iD, Youngho Seo iD, Gyuseong Cho, Javier Caravaca iD

IEEE Transactions on Nuclear Science · 2026-04 · 73 (4) · 1481-1494

DOI: 10.1109/tns.2026.3663840 ↗

Kurz gefasst

Fachartikel in der Zeitschrift IEEE Transactions on Nuclear Science, 2026-04. Verfasst von Gyohyeok Song und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Nuclear and Quantum Engineering, Korea Advanced Institute of Science and Technology, Daejeon 34141, Republic of Korea, and also with th…. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 56 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIBIB NIH HHS, National Institutes of Health, Korea Institute of Energy Technology Evaluation and Planning and the Ministry of Trade, Industry and Energy, Republic of Korea.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

We evaluated a 3-D-positioning Compton camera based on cadmium-zinc-telluride (CZT) detectors in the context of imaging of 225Ac. 225Ac, an alpha-emitting radionuclide, shows strong potential for targeted alpha therapy (TAT) of cancer. Its decay chain includes emission of 440-keV gamma rays (from 213Bi), which enables Compton imaging. Due to the extremely low injected activities of TAT radiopharmaceuticals, especially for 225Ac-based ones, the ability to measure gamma rays with high sensitivity is critical, and Compton cameras can provide higher sensitivities than collimated cameras. In Compton cameras, precise angular resolution is essential to provide good spatial resolution for accurate localization of radiopharmaceuticals. For this purpose, we customized the GDS-100 CZT camera (IDEAS) with two layers of CZT crystals to compare single- and double-layer Compton imaging. We measured and compared the key performance metrics for imaging of 225Ac, sensitivity, and angular resolution, to determine the performance of single- and double-layer CZT cameras for imaging of 225Ac. For comparison, we also characterized the camera at higher energies with a 137Cs source that emits 662-keV gamma rays. We optimized the depth of interaction (DOI) and center-of-gravity methods for 3-D positioning, followed by 3-D voxel-based energy calibration. Our results show that a double-layer configuration can improve the angular resolution measure (ARM) more than a factor of three from 14.0° to 4.3° for 137Cs (662 keV), and from 15.8° to 8.9° for 225Ac (440 keV). On the other hand, we measured a sensitivity approximately 40 times higher for the single-layer setup with respect to the double-layer setup. This is due to the small size and absorption of scattered photons in the scatterer layer. A larger double-layer Compton camera implementing more detector area is expected to provide an equivalent sensitivity with the advantage of a much better angular resolution.

Schlagwörter

Stichwörter der Autor:innen

225AcCompton imagingIndex Terms—angular resolution measure (ARM)cadmium–zinc–telluride (CZT) detector characterizationsingle- versus double-layer comparison

Autor:innen und Institutionen

  • Department of Nuclear and Quantum Engineering, Korea Advanced Institute of Science and Technology, Daejeon 34141, Republic of Korea, and also with the Department of Radiology and Biomedical imaging, University of Califo…
  • Department of Radiology and Biomedical imaging, University of California at San Francisco, San Francisco, CA 94115 USA, and also with the Nuclear Science Division, Lawrence Berkeley National Laboratory, Berkeley, CA 947…
  • Department of Radiology and Biomedical Imaging, University of California at San Francisco, San Francisco, CA 94115 USA, also with the University of California, Berkeley, CA 94720 USA, and also with the Molecular Biophys…
  • Department of Nuclear and Quantum Engineering, Korea Advanced Institute of Science and Technology, Daejeon 34141, Republic of Korea.

ORCID-Profile: Gyohyeok Song, Biswajit Das, Youngho Seo, Javier Caravaca

Förderung

  • NIBIB NIH HHS
  • National Institutes of Health
  • Korea Institute of Energy Technology Evaluation and Planning and the Ministry of Trade, Industry and Energy, Republic of Korea

Zitieren

APA

Song, G., Das, B., Seo, Y., Cho, G., & Caravaca, J. (2026). Performance Evaluation of Single- and Double-Layer CZT Compton Cameras for 225 Ac Imaging. IEEE Transactions on Nuclear Science, 73(4), 1481-1494. https://doi.org/10.1109/tns.2026.3663840

Vancouver

Song G, Das B, Seo Y, Cho G, Caravaca J. Performance Evaluation of Single- and Double-Layer CZT Compton Cameras for 225 Ac Imaging. IEEE Transactions on Nuclear Science. 2026;73(4):1481-1494. doi:10.1109/tns.2026.3663840

BibTeX
@article{Song2026,
  author = {Song, Gyohyeok and Das, Biswajit and Seo, Youngho and Cho, Gyuseong and Caravaca, Javier},
  title = {Performance Evaluation of Single- and Double-Layer CZT Compton Cameras for 225 Ac Imaging},
  journal = {IEEE Transactions on Nuclear Science},
  year = {2026},
  volume = {73},
  number = {4},
  pages = {1481--1494},
  doi = {10.1109/tns.2026.3663840}
}
Kostenlos starten →