A Pneumatically Actuated Robotic Assistant for MRI-Guided Stereotactic Neurosurgery
IEEE Transactions on Medical Robotics and Bionics · 2026-05 · 8 (2) · 906-917
DOI: 10.1109/tmrb.2026.3680983 ↗Kurz gefasst
Fachartikel in der Zeitschrift IEEE Transactions on Medical Robotics and Bionics, 2026-05. Verfasst von Gang Li und 8 weiteren Autor:innen. Beteiligte Einrichtung: Gang Li, Atharva Paralikar, Pavel Yarmolenko, Ayush Nankani, Kevin Cleary, Chima Oluigbo, and Reza Monfaredi are with Children's National Hospital, W…. Die Arbeit wurde bisher 1-mal zitiert (Crossref). Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIBIB NIH HHS, National Institutes of Health.
Abstract (Originalsprache)
This paper presents the development and cadaveric evaluation of a pneumatically actuated robotic assistant for MRI-guided stereotactic neurosurgery. The robot features two coaxial arcs that provide four degrees of freedom to position and orient a needle within an MRI scanner. It is driven by custom-designed MR-Safe pneumatic stepper motors and powered by in-room compressed medical air, enabling precise interventions under intraoperative MRI guidance. To improve anatomical visualization, two custom MR imaging coils were developed and integrated with the system. A dedicated clinical workflow was also proposed for robot-assisted stereotactic procedures under MRI-guidance. Targeting accuracy was first assessed in free space using an optical tracking system, yielding mean absolute errors of 0.97 ± 0.42 mm for tip position and 0.46 ± 0.23° for needle angle. System-level accuracy was then evaluated in MRI-guided phantom studies, showing errors of 1.38 ± 0.80 mm in tip position and 1.08 ± 0.67° in angle. Finally, a cadaver study validated the clinical workflow, with observed errors of 2.11 mm in position and 3.32° in angle. Quantitative imaging analysis confirmed minimal image degradation, with signal-to-noise ratio (SNR) variation below 12.26% and 2D geometric distortion under 0.20%.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Gang Li, Atharva Paralikar, Pavel Yarmolenko, Ayush Nankani, Kevin Cleary, Chima Oluigbo, and Reza Monfaredi are with Children's National Hospital, Washington, DC, USA.
- Wolfgang Loew and Charles Dumoulin are with Cincinnati Children's Hospital Medical Center, Cincinnati, OH, USA.
Förderung
- NIBIB NIH HHS
- National Institutes of Health
Zitieren
APA
Li, G., Paralikar, A., Yarmolenko, P., Nankani, A., Loew, W., Dumoulin, C., Cleary, K., Oluigbo, C., & Monfaredi, R. (2026). A Pneumatically Actuated Robotic Assistant for MRI-Guided Stereotactic Neurosurgery. IEEE Transactions on Medical Robotics and Bionics, 8(2), 906-917. https://doi.org/10.1109/tmrb.2026.3680983
Vancouver
Li G, Paralikar A, Yarmolenko P, Nankani A, Loew W, Dumoulin C, et al. A Pneumatically Actuated Robotic Assistant for MRI-Guided Stereotactic Neurosurgery. IEEE Transactions on Medical Robotics and Bionics. 2026;8(2):906-917. doi:10.1109/tmrb.2026.3680983
BibTeX
@article{Li2026,
author = {Li, Gang and Paralikar, Atharva and Yarmolenko, Pavel and Nankani, Ayush and Loew, Wolfgang and Dumoulin, Charles and Cleary, Kevin and Oluigbo, Chima and Monfaredi, Reza},
title = {A Pneumatically Actuated Robotic Assistant for MRI-Guided Stereotactic Neurosurgery},
journal = {IEEE Transactions on Medical Robotics and Bionics},
year = {2026},
volume = {8},
number = {2},
pages = {906--917},
doi = {10.1109/tmrb.2026.3680983}
}
EUVIMED