An Integrated Calibration Method for DC Current Transducers With I – V Circuits Using Precision DC Resistance Bridge Technology
IEEE Transactions on Instrumentation and Measurement · 2026 · 75 · 9534311-9534311
DOI: 10.1109/tim.2026.3709403 ↗Kurz gefasst
Fachartikel in der Zeitschrift IEEE Transactions on Instrumentation and Measurement, 2026. Verfasst von Qi Zhang und 9 weiteren Autor:innen. Die Arbeit nennt 19 Literaturangaben.
Abstract (Originalsprache)
A DC current transducer (DCCT) is a key component in precision DC power supplies for accelerators, where its performance, combined with downstream current-to-voltage ( $I$ – $V$ ) circuits, determines overall accuracy. Conventional calibration treats the DCCT and $I$ – $V$ circuit separately, which may overlook interaction effects such as impedance mismatches, loading effects, and other interactions, leading to system-level errors. This study proposes an integrated calibration method for DCCT measurement systems, where the combined DCCT with $I$ – $V$ circuit (DCCT-IV) is treated as an equivalent resistor and measured using a precision DC resistance bridge. The equivalent voltage corresponding to each sample current is then calculated from the measured resistance, and a fitting curve is established between the equivalent voltage and the ideal voltage, allowing the actual output voltage to approach the ideal value. Theoretical analysis and experimental results demonstrate that the proposed method effectively enhances measurement accuracy.
Zitieren
APA
Zhang, Q., Ye, Y., Han, C., Long, F., Gao, Y., Wei, P., Liu, Y., Chen, S., Guo, X., & Du, Y. (2026). An Integrated Calibration Method for DC Current Transducers With I – V Circuits Using Precision DC Resistance Bridge Technology. IEEE Transactions on Instrumentation and Measurement, 75, 9534311-9534311. https://doi.org/10.1109/tim.2026.3709403
Vancouver
Zhang Q, Ye Y, Han C, Long F, Gao Y, Wei P, et al. An Integrated Calibration Method for DC Current Transducers With I – V Circuits Using Precision DC Resistance Bridge Technology. IEEE Transactions on Instrumentation and Measurement. 2026;75:9534311-9534311. doi:10.1109/tim.2026.3709403
BibTeX
@article{Zhang2026,
author = {Zhang, Qi and Ye, Yong and Han, Chao and Long, Fengli and Gao, Yao and Wei, Pengfei and Liu, Yuting and Chen, Suying and Guo, Xiaoling and Du, Yunze},
title = {An Integrated Calibration Method for DC Current Transducers With I – V Circuits Using Precision DC Resistance Bridge Technology},
journal = {IEEE Transactions on Instrumentation and Measurement},
year = {2026},
volume = {75},
pages = {9534311--9534311},
doi = {10.1109/tim.2026.3709403}
}
EUVIMED