A Radiological Clip Design Using Ultrasound Identification to Improve Localization

Jenna Cario, Zhengchang Kou, Rita J Miller, April Dickenson, Christine U Lee, Michael L Oelze

IEEE Transactions on Biomedical Engineering · 2024

DOI: 10.1109/tbme.2024.3388203 ↗

Kurz gefasst

Fachartikel in der Zeitschrift IEEE Transactions on Biomedical Engineering, 2024. Verfasst von Jenna Cario und 5 weiteren Autor:innen. Hauptthemen laut MEDLINE: Phantoms, Imaging, Ultrasonography, Equipment Design. Die Arbeit nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS, NIBIB NIH HHS, National Institutes of Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVE: We demonstrate the use of ultrasound to receive an acoustic signal transmitted from a radiological clip designed from a custom circuit. This signal encodes an identification number and is localized and identified wirelessly by the ultrasound imaging system. METHODS: We designed and constructed the test platform with a Teensy 4.0 microcontroller core to detect ultrasonic imaging pulses received by a transducer embedded in a phantom, which acted as the radiological clip. Ultrasound identification (USID) signals were generated and transmitted as a result. The phantom and clip were imaged using an ultrasonic array (Philips L7-4) connected to a Verasonics™ Vantage 128 system operating in pulse inversion (PI) mode. Cross-correlations were performed to localize and identify the code sequences in the PI images. RESULTS: USID signals were detected and visualized on B-mode images of the phantoms with up to sub-millimeter localization accuracy. The average detection rate across 30,400 frames of ultrasound data was 98.1%. CONCLUSION: The USID clip produced identifiable, distinguishable, and localizable signals when imaged. SIGNIFICANCE: Radiological clips are used to mark breast cancer being treated by neoadjuvant chemotherapy (NAC) via implant in or near treated lesions. As NAC progresses, available marking clips can lose visibility in ultrasound, the imaging modality of choice for monitoring NAC-treated lesions. By transmitting an active signal, more accurate and reliable ultrasound localization of these clips could be achieved and multiple clips with different ID values could be imaged in the same field of view.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingUltrasonographyEquipment DesignSignal Processing, Computer-AssistedHumansTransducers

Förderung

  • NCI NIH HHS
  • NIBIB NIH HHS
  • National Institutes of Health
  • Cancer Scholars for Translational and Applied Research
  • Illinois and the Carle Cancer Center
  • Beckman Institute Graduate Fellowship
  • Beckman Institute Postdoctoral Fellowship

Zitieren

APA

Cario, J., Kou, Z., Miller, R. J., Dickenson, A., Lee, C. U., & Oelze, M. L. (2024). A Radiological Clip Design Using Ultrasound Identification to Improve Localization. IEEE Transactions on Biomedical Engineering. https://doi.org/10.1109/tbme.2024.3388203

Vancouver

Cario J, Kou Z, Miller RJ, Dickenson A, Lee CU, Oelze ML. A Radiological Clip Design Using Ultrasound Identification to Improve Localization. IEEE Transactions on Biomedical Engineering. 2024. doi:10.1109/tbme.2024.3388203

BibTeX
@article{Cario2024,
  author = {Cario, Jenna and Kou, Zhengchang and Miller, Rita J and Dickenson, April and Lee, Christine U and Oelze, Michael L},
  title = {A Radiological Clip Design Using Ultrasound Identification to Improve Localization},
  journal = {IEEE Transactions on Biomedical Engineering},
  year = {2024},
  doi = {10.1109/tbme.2024.3388203}
}
Kostenlos starten →