EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

Continuous Patient-Independent Estimation of Respiratory Rate and Blood Pressure Using Robust Spectro-Temporal Features Derived From Photoplethysmogram Only

Muhammad Ahmad Sultan, Wala Saadeh iD

IEEE Open Journal of Engineering in Medicine and Biology · 2024 · 5 · 637-649

DOI: 10.1109/ojemb.2023.3329728 ↗

Kurz gefasst

Fachartikel in der Zeitschrift IEEE Open Journal of Engineering in Medicine and Biology, 2024. Verfasst von Muhammad Ahmad Sultan und 1 weiteren Autor:innen. Beteiligte Einrichtung: Electrical Engineering DepartmentLahore University of Management Sciences (LUMS) Lahore 54792 Pakistan. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Engineering and Design Department, Western Washington University, Islamic Development Bank.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

A patient-independent approach for continuous estimation of vital signs using robust spectro-temporal features derived from only photoplethysmogram (PPG) signal.

Methods

In the pre-processing stage, we remove baseline shifts and artifacts of the PPG signal using Incremental Merge Segmentation with adaptive thresholding. From the cleaned PPG, we extract multiple parameters independent of individual patient PPG morphology for both Respiration Rate (RR) and Blood Pressure (BP). In addition, we derived a set of novel spectral and statistical features strongly correlated to BP. We proposed robust correlation-based feature selection methods for accurate RR estimates. For fewer computations and accurate measurements of BP, the most significant features are selected using correlation and mutual information measures in the feature engineering part. Finally, RR and BP are estimated using breath counting and a neural network regression model, respectively.

Results

The proposed approach outperforms the current state-of-the-art in both RR and BP. The RR algorithm results in mean absolute errors (median, 25th-75th percentiles) of 0.4 (0.1-0.7) for CapnoBase dataset and 0.5(0.3-2.8) for BIDMC dataset without discarding any data window. Similarly, BP approach has been validated on a large dataset derived from MIMIC-II ([Formula: see text]1700 records) which has errors (mean absolute, standard deviation) of 5.0(6.3) and 3.0(4.0) for systolic and diastolic BP, respectively. The results meet the American Association for the Advancement of Medical Instrumentation (AAMI) and British Hypertension Society (BHS) Class A criteria.

Conclusion

By using robust features and feature selection methods, we alleviated patient dependency to have reliable estimates of vitals.

Schlagwörter

Stichwörter der Autor:innen

Blood Pressure (BP)and wearable sensingminimal redundancy maximal relevance (mRMR)photoplethysmogram (PPG)respiration rate (RR)signal qualityvitals

Autor:innen und Institutionen

  • Electrical Engineering DepartmentLahore University of Management Sciences (LUMS) Lahore 54792 Pakistan.
  • The Engineering and Design DepartmentWestern Washington University (WWU) Bellingham WA 98225 USA.

ORCID-Profile: Wala Saadeh

Förderung

  • Engineering and Design Department
  • Western Washington University
  • Islamic Development Bank

Zitieren

APA

Sultan, M. A., & Saadeh, W. (2024). Continuous Patient-Independent Estimation of Respiratory Rate and Blood Pressure Using Robust Spectro-Temporal Features Derived From Photoplethysmogram Only. IEEE Open Journal of Engineering in Medicine and Biology, 5, 637-649. https://doi.org/10.1109/ojemb.2023.3329728

Vancouver

Sultan MA, Saadeh W. Continuous Patient-Independent Estimation of Respiratory Rate and Blood Pressure Using Robust Spectro-Temporal Features Derived From Photoplethysmogram Only. IEEE Open Journal of Engineering in Medicine and Biology. 2024;5:637-649. doi:10.1109/ojemb.2023.3329728

BibTeX
@article{Sultan2024,
  author = {Sultan, Muhammad Ahmad and Saadeh, Wala},
  title = {Continuous Patient-Independent Estimation of Respiratory Rate and Blood Pressure Using Robust Spectro-Temporal Features Derived From Photoplethysmogram Only},
  journal = {IEEE Open Journal of Engineering in Medicine and Biology},
  year = {2024},
  volume = {5},
  pages = {637--649},
  doi = {10.1109/ojemb.2023.3329728}
}
Kostenlos starten →