EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Piezoelectric Flexural Plate Wave MEMS Temperature Sensor Coupled to a Frequency Doublet Resonance Detection Circuit

Stefano Bertelli, Alessandro Nastro, Marco Ferrari, Vittorio Ferrari

IEEE Sensors Letters · 2026-12 · 10 (12) · 2505904-2505904

DOI: 10.1109/lsens.2026.3744151 ↗

Kurz gefasst

Fachartikel in der Zeitschrift IEEE Sensors Letters, 2026-12. Verfasst von Stefano Bertelli und 3 weiteren Autor:innen. Die Arbeit nennt 15 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This letter presents a piezoelectric flexural plate wave (FPW) micro electro-mechanical systems (MEMS) temperature sensor coupled to a tailored analog front-end circuit implementing the frequency doublet resonance detection (FDRD) technique. The MEMS device consists of a 6 × 6 mm composite diaphragm made of silicon and aluminum nitride layers, with metal interdigital transducers (IDTs) exciting FPWs in the first antisymmetric Lamb mode (A0). Temperature variations affect the FPW propagation velocity and, consequently, the resonance, enabling single-port temperature sensing. The FDRD method excites the IDT with a swept frequency doublet formed by two sinusoids with variable frequencies separated by a fixed gap Δf and evaluates the conductance difference ΔG at the two frequencies. The zero crossing of ΔG occurs when the doublet center frequency coincides with the resonance, allowing its detection. The sensor was characterized in a PID-controlled thermal box from 25 to 50°C, while sweeping the doublet center frequency from 12.1 to 12.2 MHz for different Δf values and the resonant frequency was detected. The measured resonance frequency showed a linear dependence on temperature, with a sensitivity of −425 Hz/°C for Δf = 5 kHz. Comparable results have been obtained for different Δf confirming that the proposed MEMS temperature sensor can be successfully coupled with the FDRD circuit over the investigated temperature range.

Zitieren

APA

Bertelli, S., Nastro, A., Ferrari, M., & Ferrari, V. (2026). Piezoelectric Flexural Plate Wave MEMS Temperature Sensor Coupled to a Frequency Doublet Resonance Detection Circuit. IEEE Sensors Letters, 10(12), 2505904-2505904. https://doi.org/10.1109/lsens.2026.3744151

Vancouver

Bertelli S, Nastro A, Ferrari M, Ferrari V. Piezoelectric Flexural Plate Wave MEMS Temperature Sensor Coupled to a Frequency Doublet Resonance Detection Circuit. IEEE Sensors Letters. 2026;10(12):2505904-2505904. doi:10.1109/lsens.2026.3744151

BibTeX
@article{Bertelli2026,
  author = {Bertelli, Stefano and Nastro, Alessandro and Ferrari, Marco and Ferrari, Vittorio},
  title = {Piezoelectric Flexural Plate Wave MEMS Temperature Sensor Coupled to a Frequency Doublet Resonance Detection Circuit},
  journal = {IEEE Sensors Letters},
  year = {2026},
  volume = {10},
  number = {12},
  pages = {2505904--2505904},
  doi = {10.1109/lsens.2026.3744151}
}