A Graphene-Based Flexible Capacitive Respiratory Sensor With CMOS Readout Circuit
IEEE Sensors Journal · 2026-08 · 26 (15) · 22248-22255
DOI: 10.1109/jsen.2026.3706498 ↗Kurz gefasst
Fachartikel in der Zeitschrift IEEE Sensors Journal, 2026-08. Verfasst von Yong-Teng Liu und 3 weiteren Autor:innen. Die Arbeit nennt 26 Literaturangaben. Gefördert von: National Science and Technology Council and the Ministry of Education, Taiwan,.
Abstract (Originalsprache)
This study demonstrates a flexible capacitive humidity and respiratory sensor achieved by integrating graphene as a sensing electrode with a polyimide-based flexible printed circuit board (FPCB). The sensing principle is governed by rapid surface adsorption and interfacial polarization, where moisture interaction effectively modulates the composite dielectric permittivity within the high-intensity electric field of the parallel-plate sensor. A dedicated CMOS readout circuit is implemented to suppress parasitic capacitance effects, thereby facilitating high-sensitivity signal acquisition. A wet-transfer process is implemented to integrate graphene onto the FPCB substrate. For relative humidity (RH) sensing, the proposed sensor exhibits a sensitivity of 8.6 fF/%RH at $25~^{\circ }$ C, with a 2.25-fold increase in capacitance as RH varies from 40%RH to 90%RH while achieving a humidity resolution of 0.12%RH. For respiratory monitoring, the sensor demonstrates a flow-rate sensitivity of 0.5 mV/liters per minute (LPM) for nasal exhalation, with a flow-rate resolution of 0.69 LPM. The rapid subsecond response of the capacitive sensing mechanism enables real-time breathing detection. The proposed graphene-based flexible sensor, combined with the CMOS readout circuit, provides a compact solution suitable for wearable and continuous respiratory monitoring applications.
Förderung
- National Science and Technology Council and the Ministry of Education, Taiwan,
Zitieren
APA
Liu, Y., Tseng, S., Yeh, C., & Lu, M. S. (2026). A Graphene-Based Flexible Capacitive Respiratory Sensor With CMOS Readout Circuit. IEEE Sensors Journal, 26(15), 22248-22255. https://doi.org/10.1109/jsen.2026.3706498
Vancouver
Liu Y, Tseng S, Yeh C, Lu MS. A Graphene-Based Flexible Capacitive Respiratory Sensor With CMOS Readout Circuit. IEEE Sensors Journal. 2026;26(15):22248-22255. doi:10.1109/jsen.2026.3706498
BibTeX
@article{Liu2026,
author = {Liu, Yong-Teng and Tseng, Sheng-Hsiang and Yeh, Chao-Hui and Lu, Michael S.-C.},
title = {A Graphene-Based Flexible Capacitive Respiratory Sensor With CMOS Readout Circuit},
journal = {IEEE Sensors Journal},
year = {2026},
volume = {26},
number = {15},
pages = {22248--22255},
doi = {10.1109/jsen.2026.3706498}
}