Vollständiger Abstract
Worum geht es in dieser Arbeit?
The increasing use of implantable neuromodulation devices in patients with chronic post-stroke sequelae raises the possibility of unintended device-device interactions. We report a case of recurrent intrathecal baclofen pump motor stalls in a patient with longstanding spasticity following an intracranial hemorrhage, ultimately attributed to magnetic interference from a recently implanted vagus nerve stimulator (VNS) activator magnet. Recognition of this interaction led to prompt resolution and highlights the importance of interdisciplinary awareness and patient education in individuals with multiple implanted neuromodulation systems.
Abstract: PubMed · Datensatz
Bibliografischer Nachweis
Publikationsdaten
- Autor:innen
- Farnoosh Naderkhani
- Quelle
- IEEE Reliability Magazine
- Publikation
- 2024-01-01
- Band / Ausgabe
- Nicht angegeben
- Seiten
- Nicht angegeben
- ISSN / ISBN
- 2641-8819
- Zitationen
- 0 laut Crossref
- Referenzen
- 0 hinterlegt
Zitieren
Zitierfähiger Nachweis
Farnoosh Naderkhani (2024). PHM. IEEE Reliability Magazine. https://doi.org/10.1097/phm.0000000000003135
Kontext