EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Validating Cleanliness of Loaned Surgical Instruments Using Adenosine Triphosphate Technology as an Adjunct to Visual Inspection

J. Michael Hartmann, Mercedes H. Maisel, Kyle C. Kinas, Prospero C. Donan, Gina L. Eberhardt

Excellence and Credentialing in Health Care · 2027-01 · 2 (1) · e0042

DOI: 10.1097/echc.0000000000000042 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Excellence and Credentialing in Health Care, 2027-01. Verfasst von J. Michael Hartmann und 4 weiteren Autor:innen. Die Arbeit nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Visual inspection has historically been the standard to validate surgical instrument cleanliness. However, this method cannot detect microscopic debris. Complex instruments, particularly loaned surgical instruments (LSI), are prone to retaining organic debris. Adenosine triphosphate (ATP) technology detects residual organic material after decontamination. This evidence-based practice project aimed to determine whether ATP combined with visual inspection improved the detection of residual organic material before and after decontamination.

Methods

From March through September 2025, inbound LSI sets were visually inspected and tested using ATP swabs before and immediately after decontamination at our facility. Sterile processing technicians performing decontamination were blinded to which instruments were tested. Visual inspections were recorded on a scale from 1 (no visible soil) to 3 (highly soiled). ATP levels were recorded in relative light units (RLU), with values ≤150 RLU considered passing and values >150 RLU considered failing.

Results

Investigators tested 72 LSI sets. Of these, 53% ( n = 38) failed initial ATP testing with RLU values >150. Of the 38 sets that failed ATP testing, 87% ( n = 33) passed visual inspection, receiving a score of 1 (no visible soil). After decontamination at the facility, all 72 instrument sets passed ATP testing (≤150 RLU) and visual inspection.

Conclusions

Visual inspection alone may be insufficient to validate the cleanliness of LSI sets. Decontamination at a previous facility does not ensure that instruments are free of residual organic material. ATP testing provides a rapid, objective, and quantifiable adjunct to visual inspection for verifying instrument cleanliness after decontamination.

Zitieren

APA

Hartmann, J. M., Maisel, M. H., Kinas, K. C., Donan, P. C., & Eberhardt, G. L. (2027). Validating Cleanliness of Loaned Surgical Instruments Using Adenosine Triphosphate Technology as an Adjunct to Visual Inspection. Excellence and Credentialing in Health Care, 2(1), e0042. https://doi.org/10.1097/echc.0000000000000042

Vancouver

Hartmann JM, Maisel MH, Kinas KC, Donan PC, Eberhardt GL. Validating Cleanliness of Loaned Surgical Instruments Using Adenosine Triphosphate Technology as an Adjunct to Visual Inspection. Excellence and Credentialing in Health Care. 2027;2(1):e0042. doi:10.1097/echc.0000000000000042

BibTeX
@article{Hartmann2027,
  author = {Hartmann, J. Michael and Maisel, Mercedes H. and Kinas, Kyle C. and Donan, Prospero C. and Eberhardt, Gina L.},
  title = {Validating Cleanliness of Loaned Surgical Instruments Using Adenosine Triphosphate Technology as an Adjunct to Visual Inspection},
  journal = {Excellence and Credentialing in Health Care},
  year = {2027},
  volume = {2},
  number = {1},
  pages = {e0042},
  doi = {10.1097/echc.0000000000000042}
}
Kostenlos starten →