EUVIMEDEUVIMED entdecken
Crossref

Empirical and semi-empirical methods to determine the kV exponent ( n value) for diagnostic X-ray machines

Wellen Rukundo iD, Natharius Nimbashabira, Samuel Wattisa Maberi, Christine Edna Emong

Radiation Protection Dosimetry · 13.11.2026

DOI: 10.1093/rpd/ncag131 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Radiation Protection Dosimetry, online erschienen am 13.11.2026. Verfasst von Wellen Rukundo und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Regulatory Affairs, Atomic Energy Council, Plot 40, Bukoto Street, P.O. Box 7044, Kampala, Uganda. Die Arbeit nennt 21 Literaturangaben.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The air kerma, and beam quality measurements were performed on 43 diagnostic X-ray machines (25 fixed and 18 mobile) that met acceptance criteria for tube potential accuracy, half-value layer, and tube output. Measurements were conducted at tube potentials from 40 kV to 120 kV using a calibrated solid-state dosimetry system at a fixed focus-detector distance of 1 m, normalized output (mGy/mAs) and half-value layer were determined at each setting of the tube potential. Two methods (method 1 – empirical and method 2 – semi-empirical) were used to derive the exponent value. In method 1, a power-law fit of tube output versus tube potential yielded an average exponent of $n$ ≈ 2.334 ± 0.118 based on the group of X-ray machines with high-power frequency generators, consistent with values reported for diagnostic systems with standard filtration [1, 2]. In method 2, the $n$ was derived as a function of tube potential by incorporating the measured variation of half-value layer with tube potential, explicitly accounting for effective filtration. This approach showed that $n$ decreases from ~2.69 at 40 kV to 2.09 at 120 kV, demonstrating greater output sensitivity to tube potential changes at lower tube potentials. Therefore, the study introduces a semi-empirical method for estimating the exponent ($n$ value) that incorporates measured half-value layer variation to account for effective beam hardening, which is neglected in conventional empirical models. This provides a practical tool for characterizing the dependence of X-ray tube output on tube protocol optimization and equipment performance assessment, applicable for quality control in diagnostic X-ray machines.

Autor:innen und Institutionen

  • Department of Regulatory Affairs, Atomic Energy Council, Plot 40, Bukoto Street, P.O. Box 7044, Kampala, Uganda.
  • Department of Physics, School of Physical Sciences, Makerere University, P.O. Box 7062, Kampala, Uganda.

ORCID-Profile: Wellen Rukundo

Zitieren

APA

Rukundo, W., Nimbashabira, N., Maberi, S. W., & Emong, C. E. (2026). Empirical and semi-empirical methods to determine the kV exponent ( n value) for diagnostic X-ray machines. Radiation Protection Dosimetry. https://doi.org/10.1093/rpd/ncag131

Vancouver

Rukundo W, Nimbashabira N, Maberi SW, Emong CE. Empirical and semi-empirical methods to determine the kV exponent ( n value) for diagnostic X-ray machines. Radiation Protection Dosimetry. 2026. doi:10.1093/rpd/ncag131

BibTeX
@article{Rukundo2026,
  author = {Rukundo, Wellen and Nimbashabira, Natharius and Maberi, Samuel Wattisa and Emong, Christine Edna},
  title = {Empirical and semi-empirical methods to determine the kV exponent ( n value) for diagnostic X-ray machines},
  journal = {Radiation Protection Dosimetry},
  year = {2026},
  doi = {10.1093/rpd/ncag131}
}
Kostenlos starten →