Sensitivity analysis of neutron and photon equivalent doses in patient organs during 3D conformal radiotherapy using Monte Carlo simulations

Asma Alem-Bezoubiri iD, Fethi Bezoubiri, Michael Speiser, Hossam Donya

Radiation Protection Dosimetry · 03.06.2026 · 202 (9) · 755-761

DOI: 10.1093/rpd/ncag027 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Radiation Protection Dosimetry, online erschienen am 03.06.2026. Verfasst von Asma Alem-Bezoubiri und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiological and Atomic Physics, Division of Physics, Nuclear Research Centre of Algiers (CRNA), B.P. 399, Alger-Gare, Algiers 16000, A…. Hauptthemen laut MEDLINE: Photons, Radiotherapy, Conformal, Neutrons, Prostatic Neoplasms. Die Arbeit nennt 12 Literaturangaben. Gefördert von: Algiers Nuclear Research Center.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study evaluates neutron and photon equivalent dose sensitivity in patient organs during prostate three-dimensional conformal radiotherapy (3DCRT) using Monte Carlo simulations. A validated Varian Clinac 2100C LINAC model and a hybrid XCAT adult male phantom were used with an 18 MV photon beam. Component-wise sensitivity analysis showed that removal of the flattening filter caused the largest increase in neutron doses, with ratios of 1.49 in the testes, 1.39 in the bladder, and 1.38 in the colon, indicating a dominant shielding role. The primary collimator also showed significant shielding, with neutron dose ratios up to 1.29 in the rectum. Treatment room shielding contributed up to 61% of the neutron dose in the brain via room-return effects. Photon doses decreased with increasing distance from the field, highlighting spatial dependence and implications for radiological protection.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

PhotonsRadiotherapy, ConformalNeutronsProstatic NeoplasmsRadiation ProtectionRadiotherapy Planning, Computer-AssistedOrgans at RiskMonte Carlo MethodHumansMaleRadiotherapy DosagePhantoms, ImagingComputer SimulationUrinary BladderRadiometry

Autor:innen und Institutionen

  • Department of Radiological and Atomic Physics, Division of Physics, Nuclear Research Centre of Algiers (CRNA), B.P. 399, Alger-Gare, Algiers 16000, Algeria.
  • Atomic Energy Commission (COMENA), 02, Boulevard Frantz Fanon, Algiers, Algeria.
  • Englewood Hospital and Medical Center, Englewood, NJ, United States.
  • Department of Physics, Faculty of Science, King Abdulaziz University, Jeddah 21589, Saudi Arabia.
  • Department of Physics, Faculty of Science, Menoufia University, Shibin El-Koom 32511, Egypt.

ORCID-Profile: Asma Alem-Bezoubiri

Förderung

  • Algiers Nuclear Research Center

Zitieren

APA

Alem-Bezoubiri, A., Bezoubiri, F., Speiser, M., & Donya, H. (2026). Sensitivity analysis of neutron and photon equivalent doses in patient organs during 3D conformal radiotherapy using Monte Carlo simulations. Radiation Protection Dosimetry, 202(9), 755-761. https://doi.org/10.1093/rpd/ncag027

Vancouver

Alem-Bezoubiri A, Bezoubiri F, Speiser M, Donya H. Sensitivity analysis of neutron and photon equivalent doses in patient organs during 3D conformal radiotherapy using Monte Carlo simulations. Radiation Protection Dosimetry. 2026;202(9):755-761. doi:10.1093/rpd/ncag027

BibTeX
@article{AlemBezoubiri2026,
  author = {Alem-Bezoubiri, Asma and Bezoubiri, Fethi and Speiser, Michael and Donya, Hossam},
  title = {Sensitivity analysis of neutron and photon equivalent doses in patient organs during 3D conformal radiotherapy using Monte Carlo simulations},
  journal = {Radiation Protection Dosimetry},
  year = {2026},
  volume = {202},
  number = {9},
  pages = {755--761},
  doi = {10.1093/rpd/ncag027}
}
Kostenlos starten →