EUVIMEDEUVIMED entdecken
Crossref

Radiation dose to the eye lens and the shielding efficacy of a protective eye mask in patients with Hodgkin lymphoma undergoing repeated CT examinations

Do-Byung Rhee iD, Jin-Seok Yeo iD, Hyo-Gi Kim iD, Il-Hwan Bae iD, Hee-June Kim iD

Radiation Protection Dosimetry · 2025

DOI: 10.1093/rpd/ncaf092 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Radiation Protection Dosimetry, 2025. Verfasst von Do-Byung Rhee und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Kyungpook National University Chilgok Hospital, Daegu 41404, South Korea. Hauptthemen laut MEDLINE: Hodgkin Disease, Lens, Crystalline, Tomography, X-Ray Computed, Radiation Protection. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 37 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study evaluated radiation dose to the eye lens and the shielding efficacy of a radiation-protective eye mask (E-Mask) during repeated chest and abdomen computed tomography (CT) scans. Dose measurements were obtained using an anthropomorphic phantom and an optically stimulated luminescence dosemeter (Model TOPAZ, LANDAUER, USA). A 5-y follow-up scenario for Hodgkin lymphoma was simulated by assuming one scan per quarter, with the single-scan dose multiplied by 20 to represent four scans annually over 5-y. Despite the use of low-dose protocols based on national diagnostic reference levels issued by the Korea Disease Control and Prevention Agency, unshielded cumulative lens doses reached 363.89 mSv (right) and 424.84 mSv (left). Using an E-Mask reduced these doses to 243.96 mSv and 240.13 mSv, corresponding to reductions of 32.96% and 43.48%, respectively. These findings support the use of the E-Mask in radiosensitive populations to reduce lens exposure during long-term CT follow-up.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Hodgkin DiseaseLens, CrystallineTomography, X-Ray ComputedRadiation ProtectionEye Protective DevicesHumansPhantoms, ImagingRadiation DosageRadiation Injuries

Autor:innen und Institutionen

  • Department of Radiology, Kyungpook National University Chilgok Hospital, Daegu 41404, South Korea.
  • Department of Biomedical Engineering, School of Medicine, Catholic University of Daegu, Daegu 42472, South Korea.
  • Department of Anesthesiology and Pain Medicine, School of Medicine, Kyungpook National University, Kyungpook National University Chilgok Hospital, Daegu 41404, South Korea.
  • Department of Orthopaedic Surgery, School of Medicine, Kyungpook National University, Kyungpook National University Hospital, Daegu 41944, South Korea.

ORCID-Profile: Do-Byung Rhee, Jin-Seok Yeo, Hyo-Gi Kim, Il-Hwan Bae, Hee-June Kim

Zitieren

APA

Rhee, D. B., Yeo, J. S., Kim, H. G., Bae, I. H., & Kim, H. J. (2025). Radiation dose to the eye lens and the shielding efficacy of a protective eye mask in patients with Hodgkin lymphoma undergoing repeated CT examinations. Radiation Protection Dosimetry. https://doi.org/10.1093/rpd/ncaf092

Vancouver

Rhee DB, Yeo JS, Kim HG, Bae IH, Kim HJ. Radiation dose to the eye lens and the shielding efficacy of a protective eye mask in patients with Hodgkin lymphoma undergoing repeated CT examinations. Radiation Protection Dosimetry. 2025. doi:10.1093/rpd/ncaf092

BibTeX
@article{Rhee2025,
  author = {Rhee, Do-Byung and Yeo, Jin-Seok and Kim, Hyo-Gi and Bae, Il-Hwan and Kim, Hee-June},
  title = {Radiation dose to the eye lens and the shielding efficacy of a protective eye mask in patients with Hodgkin lymphoma undergoing repeated CT examinations},
  journal = {Radiation Protection Dosimetry},
  year = {2025},
  doi = {10.1093/rpd/ncaf092}
}
Kostenlos starten →