Evaluation of the self-shielding characteristics of an innovative radiosurgical system ZAP-X
Radiation Protection Dosimetry · 2025
DOI: 10.1093/rpd/ncaf007 ↗Kurz gefasst
Evaluationsstudie in der Zeitschrift Radiation Protection Dosimetry, 2025. Verfasst von Li Mingsheng und 5 weiteren Autor:innen. Beteiligte Einrichtung: Key Laboratory of Radiological Protection and Radiation Emergency, National Institute for Radiological Protection, China CDC, Beijing, China. Hauptthemen laut MEDLINE: Radiation Protection, Radiosurgery, Radiation Monitoring, Occupational Exposure. Die Arbeit nennt 8 Literaturangaben.
Abstract (Originalsprache)
Purpose
The aim of this study was to evaluate the radiation level of a ZAP-X self-shielding system by comparing it with the requirements of the Chinese radiation protection standards and to provide a practical solution for the operation of ZAP-X in China.
Methods
Under the maximum workload of the ZAP-X system, the weekly radiation dose and the instantaneous dose rate (IDR) outside the safety perimeter of the ZAP-X system were compared to the requirements of relevant standards.
Results
The maximum weekly and annual dose equivalents, derived from the IDRs and accumulative doses, were as follows:1.99 μSv/week and 0.10 mSv/year, and 3.47 μSv/week and 0.17 mSv/year, respectively.
Conclusion
The radiation levels around the ZAP-X system meet the requirements of the weekly and annual dose constraints in relevant Chinese standards. The IDRs generated by the ZAP-X system do not meet the requirements of the existing radiation protection standards. A viable solution is to adjust the safety perimeter to comply with existing standards.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Key Laboratory of Radiological Protection and Radiation Emergency, National Institute for Radiological Protection, China CDC, Beijing, China.
Zitieren
APA
Mingsheng, L., Xiangjie, M., Ya, L., Xiao, W., Quanfu, S., & Jinsheng, C. (2025). Evaluation of the self-shielding characteristics of an innovative radiosurgical system ZAP-X. Radiation Protection Dosimetry. https://doi.org/10.1093/rpd/ncaf007
Vancouver
Mingsheng L, Xiangjie M, Ya L, Xiao W, Quanfu S, Jinsheng C. Evaluation of the self-shielding characteristics of an innovative radiosurgical system ZAP-X. Radiation Protection Dosimetry. 2025. doi:10.1093/rpd/ncaf007
BibTeX
@article{Mingsheng2025,
author = {Mingsheng, Li and Xiangjie, Ma and Ya, Liu and Xiao, Wei and Quanfu, Sun and Jinsheng, Cheng},
title = {Evaluation of the self-shielding characteristics of an innovative radiosurgical system ZAP-X},
journal = {Radiation Protection Dosimetry},
year = {2025},
doi = {10.1093/rpd/ncaf007}
}
EUVIMED