The effect of cooling rate on the precision of measurement of glow peak 5 in the thermoluminescence of LiF:Mg,Ti (TLD-100)
Radiation Protection Dosimetry · 2024
DOI: 10.1093/rpd/ncad318 ↗Kurz gefasst
Fachartikel in der Zeitschrift Radiation Protection Dosimetry, 2024. Verfasst von Yogev Bokobza und 9 weiteren Autor:innen. Beteiligte Einrichtung: Physics Unit, Sami Shamoon College of Engineering, Beer-Sheva 8428322, Israel. Hauptthemen laut MEDLINE: Thermoluminescent Dosimetry, Lithium Compounds. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 6 Literaturangaben.
Abstract (Originalsprache)
The effect of natural rapid cooling and oven slow cooling on the precision of thermoluminescence measurements of LiF:Mg,Ti is investigated. Three separate series of measurements resulted in average precisions of 5.1 and 5.0%, respectively. However, the highest precision of 1.7% (1 SD) was achieved for an oven-cooled material.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Physics Unit, Sami Shamoon College of Engineering, Beer-Sheva 8428322, Israel.
- Physics Department, Ben Gurion University of the Negev, Beer-Sheva 8410501, Israel.
- Soreq Nuclear Research Center, Yavne 81800, Israel.
- Nuclear Research Centre Negev, Beer-Sheva 84190, Israel.
ORCID-Profile: Leonid Oster
Zitieren
APA
Bokobza, Y., Biderman, S., Horowitz, Y. S., Oster, L., Eliyahu, I., Shapiro, A., Nemirovsky, D., Reshes, G., Herman, B., & Einav, H. (2024). The effect of cooling rate on the precision of measurement of glow peak 5 in the thermoluminescence of LiF:Mg,Ti (TLD-100). Radiation Protection Dosimetry. https://doi.org/10.1093/rpd/ncad318
Vancouver
Bokobza Y, Biderman S, Horowitz YS, Oster L, Eliyahu I, Shapiro A, et al. The effect of cooling rate on the precision of measurement of glow peak 5 in the thermoluminescence of LiF:Mg,Ti (TLD-100). Radiation Protection Dosimetry. 2024. doi:10.1093/rpd/ncad318
BibTeX
@article{Bokobza2024,
author = {Bokobza, Yogev and Biderman, Shlomo and Horowitz, Yigal Shalom and Oster, Leonid and Eliyahu, Ilan and Shapiro, Alex and Nemirovsky, Dmitry and Reshes, Galina and Herman, Beny and Einav, Hezy},
title = {The effect of cooling rate on the precision of measurement of glow peak 5 in the thermoluminescence of LiF:Mg,Ti (TLD-100)},
journal = {Radiation Protection Dosimetry},
year = {2024},
doi = {10.1093/rpd/ncad318}
}