High-count-rate effects in event processing for the XRISM/Resolve X-ray microcalorimeter. II. Energy scale and resolution in orbit

Misaki Mizumoto, Yoshiaki Kanemaru, Shinya Yamada, Caroline A Kilbourne, Megan E Eckart, Edmund Hodges-Kluck, Yoshitaka Ishisaki, Frederick S Porter, Katja Pottschmidt, Tsubasa Tamba

Publications of the Astronomical Society of Japan · 30.09.2025 · 77 (Supplement_1) · S39-S49

DOI: 10.1093/pasj/psaf071 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Publications of the Astronomical Society of Japan, online erschienen am 11.07.2025. Verfasst von Misaki Mizumoto und 9 weiteren Autor:innen. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The Resolve instrument on the X-ray Imaging and Spectroscopy Mission (XRISM) uses a 36 pixel microcalorimeter designed to deliver high-resolution, non-dispersive X-ray spectroscopy. Although it is optimized for extended sources with low count rates, Resolve observations of bright point sources are still able to provide unique insights into the physics of these objects, as long as high-count-rate effects are addressed in the analysis. These effects include the loss of exposure time for each pixel, changes in the energy scale, and changes in the energy resolution. To investigate these effects under realistic observational conditions, we observed the bright X-ray source, the Crab Nebula, with XRISM at several offset positions with respect to the Resolve field of view and with continuous illumination from 55Fe sources on the filter wheel. For the spectral analysis, we excluded data where exposure-time loss was too significant to ensure reliable spectral statistics. The energy scale at 6 keV shows a slight negative shift in the high-count-rate regime. The energy resolution at 6 keV worsens as the count rate in electrically neighboring pixels increases, but can be restored by applying a nearest-neighbor coincidence cut (“cross-talk cut”). We examined how these effects influence the observation of bright point sources, using GX 13+1 as a test case, and identified an eV-scale energy offset at 6 keV between the inner (brighter) and outer (fainter) pixels. Users who seek to analyze velocity structures on the order of tens of km s–1 should account for such high-count-rate effects. These findings will aid in the interpretation of Resolve data from bright sources and provide valuable considerations for designing and planning for future microcalorimeter missions.

Zitieren

APA

Mizumoto, M., Kanemaru, Y., Yamada, S., Kilbourne, C. A., Eckart, M. E., Hodges-Kluck, E., Ishisaki, Y., Porter, F. S., Pottschmidt, K., & Tamba, T. (2025). High-count-rate effects in event processing for the XRISM/Resolve X-ray microcalorimeter. II. Energy scale and resolution in orbit. Publications of the Astronomical Society of Japan, 77(Supplement_1), S39-S49. https://doi.org/10.1093/pasj/psaf071

Vancouver

Mizumoto M, Kanemaru Y, Yamada S, Kilbourne CA, Eckart ME, Hodges-Kluck E, et al. High-count-rate effects in event processing for the XRISM/Resolve X-ray microcalorimeter. II. Energy scale and resolution in orbit. Publications of the Astronomical Society of Japan. 2025;77(Supplement_1):S39-S49. doi:10.1093/pasj/psaf071

BibTeX
@article{Mizumoto2025,
  author = {Mizumoto, Misaki and Kanemaru, Yoshiaki and Yamada, Shinya and Kilbourne, Caroline A and Eckart, Megan E and Hodges-Kluck, Edmund and Ishisaki, Yoshitaka and Porter, Frederick S and Pottschmidt, Katja and Tamba, Tsubasa},
  title = {High-count-rate effects in event processing for the XRISM/Resolve X-ray microcalorimeter. II. Energy scale and resolution in orbit},
  journal = {Publications of the Astronomical Society of Japan},
  year = {2025},
  volume = {77},
  number = {Supplement_1},
  pages = {S39--S49},
  doi = {10.1093/pasj/psaf071}
}
Kostenlos starten →