Displacement-pressure biparametrically regulated softness sensory system for intraocular pressure monitoring
National Science Review · 03.05.2024 · 11 (6)
DOI: 10.1093/nsr/nwae050 ↗Kurz gefasst
Fachartikel in der Zeitschrift National Science Review, online erschienen am 06.02.2024. Verfasst von Yu Cheng und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Materials Science and Engineering, Southern University of Science and Technology, Shenzhen 518055, China. Die Arbeit wurde bisher 24-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Natural Science Foundation of China, Science Technology and Innovation Committee of Shenzhen Municipality, Shenzhen Sci-Tech Fund.
Abstract (Originalsprache)
High intraocular pressure (IOP) is one of the high-risk pathogenic factors of glaucoma. Existing methods of IOP measurement are based on the direct interaction with the cornea. Commercial ophthalmic tonometers based on snapshot measurements are expensive, bulky, and their operation requires trained personnel. Theranostic contact lenses are easy to use, but they may block vision and cause infection. Here, we report a sensory system for IOP assessment that uses a soft indentor with two asymmetrically deployed iontronic flexible pressure sensors to interact with the eyelid-eyeball in an eye-closed situation. Inspired by human fingertip assessment of softness, the sensory system extracts displacement-pressure information for soft evaluation, achieving high accuracy IOP monitoring (>96%). We further design and custom-make a portable and wearable ophthalmic tonometer based on the sensory system and demonstrate its high efficacy in IOP screening. This sensory system paves a way towards cost-effective, robust, and reliable IOP monitoring.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Materials Science and Engineering, Southern University of Science and Technology, Shenzhen 518055, China.
- Department of Engineering Science and Mechanics, Pennsylvania State University, State College 16802, USA.
- Department of Biomedical Engineering, Pennsylvania State University, State College 16802, USA.
- Department of Materials Science and Engineering, Materials Research Institute, Pennsylvania State University, State College 16802, USA.
ORCID-Profile: Chuan Fei Guo
Förderung
- National Natural Science Foundation of China
- Science Technology and Innovation Committee of Shenzhen Municipality
- Shenzhen Sci-Tech Fund
- Guangdong Provincial Key Laboratory Program
Zitieren
APA
Cheng, Y., Zhan, Y., Guan, F., Shi, J., Wang, J., Sun, Y., Zubair, M., Yu, C., & Guo, C. F. (2024). Displacement-pressure biparametrically regulated softness sensory system for intraocular pressure monitoring. National Science Review, 11(6). https://doi.org/10.1093/nsr/nwae050
Vancouver
Cheng Y, Zhan Y, Guan F, Shi J, Wang J, Sun Y, et al. Displacement-pressure biparametrically regulated softness sensory system for intraocular pressure monitoring. National Science Review. 2024;11(6). doi:10.1093/nsr/nwae050
BibTeX
@article{Cheng2024,
author = {Cheng, Yu and Zhan, Yifei and Guan, Fangyi and Shi, Junli and Wang, Jingxiao and Sun, Yi and Zubair, Muhammad and Yu, Cunjiang and Guo, Chuan Fei},
title = {Displacement-pressure biparametrically regulated softness sensory system for intraocular pressure monitoring},
journal = {National Science Review},
year = {2024},
volume = {11},
number = {6},
doi = {10.1093/nsr/nwae050}
}